Analysis of minocycline by high-performance liquid chromatography in tissue and serum |
| |
Authors: | WR Wrightson SR Myers S Galandiuk |
| |
Affiliation: | Department of Surgery, University of Louisville School of Medicine, KY 40292, USA. |
| |
Abstract: | ![]() A sensitive and rapid reversed-phase high-performance liquid chromatography assay can be used to accurately determine serum and tissue minocycline concentrations. Minocycline is a broad spectrum tetracycline derivative with many applications. Tissue and serum samples were obtained from guinea pigs that had received either topical or intravenous minocycline. Samples were extracted using a Sep-Pak C18 cartridge and were injected into a microBondapak C18 column with an isocratic methanol mobile phase. Samples were analyzed using UV detection and produced sharp peaks with a retention time of 2.5 min. The lower limit of detection was 100 ng and drug recovery was 61%. This method greatly facilitated the analysis of minocycline while allowing for sensitivity. |
| |
Keywords: | |
本文献已被 PubMed 等数据库收录! |
|