Amperometric glucose biosensor based on immobilization of glucose oxidase in electropolymerized o-aminophenol film at Prussian blue-modified platinum electrode
Authors:Dawei Pan  Lihua Nie  Wenyan Tao  Shouzhuo Yao
Affiliation:State Key Laboratory of Chemo/Biosensing and Chemometrics, College of Chemistry and Chemical Engineering, Hunan University, Changsha 410082, PR China
Abstract:An amperometric glucose biosensor is developed, based on immobilization of glucose oxidase (GOX) in an electrochemically polymerized, non-conducting poly(o-aminophenol) (POAP) film at Prussian blue (PB)-modified platinum (Pt) microelectrode. Effects of polymerization cycle number for POAP and PB, applied potential used in the determination, pH value of the detection solution and electroactive compounds on the amperometric response of the sensor were investigated and discussed. The electroactive property and rough surface of PB film result in the improvement of the detection limit and the increase of the maximum response current and sensitivity. The biosensor based on Pt/PB/POAP/GOX electrode has two times lower detection limit, five times larger maximum current and nine times higher sensitivity than those of the biosensor based on Pt/POAP/GOX electrode. Additionally, the biosensor shows fast response time, large response current, and good anti-interferent ability for l-ascorbic acid, uric acid and acetaminophen. Excellent reproducibility and stability of biosensor are also observed.
Keywords:Glucose oxidase  Poly(o-aminophenol)  Prussian blue  Electrochemical polymerization  Biosensor