首页 | 本学科首页   官方微博 | 高级检索  
     

立体定向放射治疗计划输出剂量的验证
引用本文:胡家成,杨小元,杨元第,朱烨.立体定向放射治疗计划输出剂量的验证[J].现代计量测试,2005(3).
作者姓名:胡家成  杨小元  杨元第  朱烨
作者单位:中国计量科学研究院 北京100013 (胡家成,杨小元,杨元第),中国计量科学研究院 北京100013(朱烨)
摘    要:本文简单介绍对加速器的立体定向治疗计划的输出剂量进行的验证测量,其测量结果和治疗计划预制输出剂量在-1 1%内吻合。并利用胶片法测量了立体定向治疗计划的输出剂量的分布,其测量结果与治疗计划输出剂量分布的80%等剂量曲线面积重合率为94%,得到较满意的结果。

关 键 词:立体定向  治疗计划  吸收剂量

The verification for transport dose of stereotactic radiation therapy plan
HU Jiacheng,YANG Xiaoyuan,YANG Yuandi,ZHU Ye.The verification for transport dose of stereotactic radiation therapy plan[J].Modern Measurement and Test,2005(3).
Authors:HU Jiacheng  YANG Xiaoyuan  YANG Yuandi  ZHU Ye
Abstract:The paper simply introduce the verify measurement for transport dose of stereotactic therapy plan system of accelerator, the result of measure and transport dose of therapy plan which prospected differ by -1 1% The distribution of stereotactic transport dose was measured by film method The overlap rate between the area of film actually which was radiated and the distribution of stereotactic transport dose in therapy plan within 94% The result is very satisfyed
Keywords:stereotactic  therapy plan  absorb dose