首页 | 本学科首页   官方微博 | 高级检索  
     


Calibration of Thermocouples and Infrared Radiation Thermometers by Comparison to Radiation Thermometry
Authors:M J Martin  M Zarco  D Del Campo
Affiliation:1. Centro Espa?ol de Metrolog??a (CEM), Tres Cantos, Spain
Abstract:The National Metrology Institute of Spain (CEM) has designed, characterized, and set-up its new system to calibrate thermocouples and infrared radiation thermometers up to 1600 °C by comparison to radiation thermometry. This system is based on a MoSi2 three-zone furnace with a graphite blackbody comparator. Two interchangeable alumina tubes with different structures are used for thermocouples and radiation thermometer calibrations. The reference temperature of the calibration is determined by a standard radiation thermometer. Normally, this is used at CEM to disseminate the International Temperature Scale of 1990 (ITS-90) in the radiation range, and it refers to the Cu fixed point. Several noble metal thermocouples and infrared radiation thermometers with a central wavelength near 900 nm have been calibrated, and their uncertainty budgets have been obtained.
Keywords:
本文献已被 SpringerLink 等数据库收录!