A vibrating tube flow densitometer for measurements with corrosive solutions at temperatures up to 723 K and pressures up to 40 MPa |
| |
Authors: | V. Hynek M. Obšil J. Quint J. -P. E. Grolier |
| |
Affiliation: | (1) Department of Physical Chemistry, Institute of Chemical Technology, Prague, Czech Republic;(2) Laboratoire de Thermodynamique et Génie Chimique, Université Blaise Pascal/CNRS, Clermont-Ferrand, France |
| |
Abstract: | A new version of a vibrating tube flow densitometer has been designed permitting measurements of density differences between two fluids in the temperature range from 298 to 723 K and at pressures up to 40 MPa. The instrument is equipped with a Pt/Rh20 vibrating tube (1.6-mm o.d.) and a Pt/Rh10 transporting tube (1.2-mm o.d.) permitting measurements with highly corrosive liquids. The period of oscillation of the tube is about 7.5 ms, with a typical stability better than 10−4% over about a 1-h period over the entire temperature interval. The calibration constantK at room temperature is about 530 kg·m−3·ms−2, with a temperature coefficient of approximately −0.13kg·m−3·ms−2·K−1, and is practically pressure independent. It can be determined by calibration with a reproducibility generally better than 0.1%. The instrument was tested with NaCl(aq) solutions in the temperature range from 373 to 690 K for density differences between sample and reference liquid ranging from 200 to 2 kg·m−3; the corresponding errors are believed to be below 0.3 and 5%, respectively. A highly automated temperature control maintains the temperature of the tube stable to within ±0.02 K. |
| |
Keywords: | aqueous solutions critical point flow densitometer high temperature high pressure vibrating tube |