离子色谱法测定饮用水中的磷酸盐
引用本文:陈少宏,吴文威.离子色谱法测定饮用水中的磷酸盐[J].福建分析测试,2013,22(2):35-37.
作者姓名:陈少宏  吴文威
作者单位:中山市环境监测站,广东中山,528400
摘    要:建立了单柱阴离子色谱测定饮用水中磷酸盐的方法,并对实际水样进行了测定。磷酸盐浓度在0.1-4.0mg/L范围内与峰面积线性回归相关系数0.9992,检出限为0.02mg/L,保留时间和峰面积相对标准偏差分别为0.09%和0.52%,回收率为96.2-102.4%。

关 键 词:离子色谱  磷酸盐  饮用水

Analysis of phosphate in drinking water Using Ion Chromatography
Chen Shao-hong , Wu Wen-wei.Analysis of phosphate in drinking water Using Ion Chromatography[J].Fujian Anal Ysis & Testing,2013,22(2):35-37.
Authors:Chen Shao-hong  Wu Wen-wei
Affiliation:(Zhongshan Environmental Monitoring Station,Zhongshan,Zhongshan,Guangdong 528400,China)
Abstract:A method for the determination of phosphate in drinking water samples by using ion chromatography was established. Linear correlation coefficient was 0.9992. The limits of detection were 0.02mg/L. The relative standard deviation of retention time and peak area were 0.09% and 0.52%. The recoveries of eight phthalic acid esters were 96.2-102.4%.
Keywords:ion chromatography  phosphate  drinking water