首页 | 本学科首页   官方微博 | 高级检索  
     


Angiojet™ mechanical thrombectomy-induced hemolysis triggering blood-leak alarms
Authors:Babitha Selvananthan  Omar Aziz  Arvind D Lee  Keagan Werner-Gibbings  Bhadran Bose  Kamal Sud
Affiliation:1. Nepean Kidney Research Centre, Department of Renal Medicine, Nepean Hospital, Kingswood, Australia;2. Department of Vascular Surgery, Nepean Hospital, Kingswood, Australia;3. Nepean Kidney Research Centre, Department of Renal Medicine, Nepean Hospital, Kingswood, Australia

Nepean Clinical School, Faculty of Medicine and Health, The University of Sydney, Kingswood, Australia

Abstract:Blood leak alarms are increasingly rare and may be triggered by rupture of the dialyzer membrane, or intravascular hemolysis. We report two patients who developed hemolysis following Angiojet™ thrombolysis and thrombectomy of thrombosed arteriovenous fistulas, triggering blood leak alarms on hemodialysis, the occurrence of which has not been reported before. AngioJet™-induced hemolysis should be considered in the differential diagnosis for blood leak alarms occurring soon after an intervention on an arterio-venous (AV) fistula.
Keywords:Angiojet™ thrombolysis and thrombectomy  blood leak alarms  hemolysis  hemodialysis
设为首页 | 免责声明 | 关于勤云 | 加入收藏