首页 | 本学科首页   官方微博 | 高级检索  
     


Errors in strain measurement at high frequency
Authors:M Dixon
Affiliation:Cambridge Consultants Ltd, Science Park, Milton Road, Cambridge CB4 4DW
Abstract:Two possible sources of error in strain measurement at high frequency were considered. The performance of the strain gauge itself was measured by comparison with a reference transducer. Several installations were shown to he satisfactory at up to 100 Hz hut another showed a progressive attenuation with frequency. Attempts to reproduce the fault with another installation were unsuccessful. The sources of error within the instrumentation were also considered and measurements made on the dynamic performance of several voltmeters.
Keywords:Dynamic strain measurement  transducer  calibration  instrumentation bandwidth  
设为首页 | 免责声明 | 关于勤云 | 加入收藏