首页 | 本学科首页   官方微博 | 高级检索  
     


High-resolution confocal microscopy using synchrotron radiation
Authors:C J R VAN DER OORD  G R JONES  D A SHAW  I H MUNRO  Y K LEVINE  & H C GERRITSEN
Affiliation:Department of Molecular Biophysics, Utrecht University, PO Box 80.000, 3508 TA Utrecht, The Netherlands;, SERC Laboratory Daresbury, Warrington WA4 4AD, U.K.
Abstract:A confocal scanning light microscope coupled to the Daresbury Synchrotron Radiation Source is described. The broad spectrum of synchrotron radiation and the application of achromatic quartz/CaF2 optics allows for confocal imaging over the wavelength range 200–700 nm. This includes UV light, which is particularly suitable for high-resolution imaging. The results of test measurements using 290-nm light indicate that a lateral resolution better than 100 nm is obtained. An additional advantage of the white synchrotron radiation is that the excitation wavelength can be chosen to match the absorption band of any fluorescent dye. The availability of UV light for confocal microscopy enables studies of naturally occurring fluorophores. The potential applications of the microscope are illustrated by the real-time imaging of hormone traffic using the naturally occurring oestrogen coumestrol. (The IUPAC name for coumestrol is 3,9-dihydroxy-6H-benzofuro3,2-c]1]benzopyran-6-one ( Chem. Abstr. Reg. No . 479-13-0). The trivial name will be used throughout this paper.)
Keywords:Confocal microscopy  synchrotron radiation  ultraviolet microscopy