A comparative analysis of the sensitivity of CsI (Tl), ZnO(Ga), and YAG(Ce) scintillators to the plasma background radiation under operating conditions of the ITER tokamak reactor |
| |
Authors: | Email author" target="_blank">V?G?NesenevichEmail author V?I?Afanasyev S?S?Kozlovskii D?V?Makar’in A?D?Mel’nik M?I?Mironov M?P?Petrov S?Ya?Petrov F?V?Chernyshev |
| |
Affiliation: | 1.Ioffe Physical Technical Institute,Russian Academy of Sciences,St. Petersburg,Russia;2.St. Petersburg State Polytechnic University,St. Petersburg,Russia |
| |
Abstract: | Problems related to neutral particle flux measurements on the ITER tokamak reactor under intense plasma background radiation
conditions are considered. The results of measuring a background sensitivity with respect to neutron and γ-radiation for the
scintillation detector, which is based on three different crystals (CsI (Tl), ZnO(Ga), and YAG(Ce)), are presented. The scintillators
are compared and conclusions about the possibility of their applications in detectors of neutral particle analyzers currently
created at the Ioffe Institute for the ITER tokamak reactor, are drawn. |
| |
Keywords: | |