首页 | 本学科首页   官方微博 | 高级检索  
     

Offner双镜三反射成像光谱仪分辨率的研究
引用本文:裴梓任,黄元申,倪争技.Offner双镜三反射成像光谱仪分辨率的研究[J].光学仪器,2014,36(2):147-151.
作者姓名:裴梓任  黄元申  倪争技
作者单位:上海理工大学 光电信息与计算机工程学院, 上海 200093;上海理工大学 光电信息与计算机工程学院, 上海 200093;上海理工大学 光电信息与计算机工程学院, 上海 200093
基金项目:国家自然科学基金资助项目(61205156、61378060);科技部重大科学仪器专项(2011YQ15004002、2011YQ15004004、2011YQ14014704);上海市教委曙光项目(11SG44)
摘    要:针对Offner双镜三反射成像光谱仪的消像差结构,采用几何方法推导出光谱分辨率的计算公式,分析了入射狭缝的宽度、凸面光栅分辨率、系统像差和探测器像元尺寸各个参数对光谱分辨率的影响,提出了分光系统像差的计算方法和优化设计方法,并探讨了提高光谱分辨率的方法和技术,即在优化系统像差的同时,适当减小狭缝宽度和探测器像元尺寸,有利于提高系统的光谱分辨率。该系统利用消像差优化设计同时考虑光谱分辨率的设计方法,具有十分重要的实用价值,为成像光谱仪的研制提供经验和借鉴。

关 键 词:光栅光谱仪  光谱分辨率  Offner结构  同心光学系统
收稿时间:2013/12/12

Research of resolution for Offner two-mirror three-reflection imaging spectrometer
PEI Ziren,HUANG Yuanshen and NI Zhengji.Research of resolution for Offner two-mirror three-reflection imaging spectrometer[J].Optical Instruments,2014,36(2):147-151.
Authors:PEI Ziren  HUANG Yuanshen and NI Zhengji
Affiliation:School of Optical-Electrical and Computer Engineering, University of Shanghai for Science and Technology, Shanghai 200093, China;School of Optical-Electrical and Computer Engineering, University of Shanghai for Science and Technology, Shanghai 200093, China;School of Optical-Electrical and Computer Engineering, University of Shanghai for Science and Technology, Shanghai 200093, China
Abstract:The calculating equation of spectral resolution was deduced by geometry method based on eliminating aberration structure of Offner two-mirror three-reflection imaging spectrometer. The entrance slit width, convex grating resolution, system aberration and pixel size of detector have a great influence on spectral resolution, and the influence of resolution for each parameter was analyzed. The calculating method and optimization design method of aberration were proposed in spectroscopic system. The techniques and methods of improving spectral resolution were summarized, namely, when the system aberration was optimized, the spectral resolution could be enhanced by decreasing entrance slit width and pixel size of detector. This system used optimization design of eliminating aberration, as well as design method of spectral resolution, which has important practical value. It can provide the experience and reference for imaging spectrometer's designing.
Keywords:grating spectrometer  spectral resolution  Offner configuration  concentric optical system
本文献已被 CNKI 维普 等数据库收录!
点击此处可从《光学仪器》浏览原始摘要信息
点击此处可从《光学仪器》下载全文