首页 | 本学科首页   官方微博 | 高级检索  
     


Defocus and twofold astigmatism correction in HAADF-STEM
Authors:Rudnaya M E  Van den Broek W  Doornbos R M P  Mattheij R M M  Maubach J M L
Affiliation:a CASA, Department of Mathematics and Computer Science, Eindhoven University of Technology, PO Box 513, 5600 MB Eindhoven, The Netherlands
b EMAT, University of Antwerp, Groenenborgerlaan 171, B-2020 Antwerp, Belgium
c Embedded Systems Institute, PO Box 513, 5600 MB Eindhoven, The Netherlands
Abstract:A new simultaneous autofocus and twofold astigmatism correction method is proposed for High Angle Annular Dark Field Scanning Transmission Electron Microscopy (HAADF-STEM). The method makes use of a modification of image variance, which has already been used before as an image quality measure for different types of microscopy, but its use is often justified on heuristic grounds. In this paper we show numerically that the variance reaches its maximum at Scherzer defocus and zero astigmatism. In order to find this maximum a simultaneous optimization of three parameters (focus, x- and y-stigmators) is necessary. This is implemented and tested on a FEI Tecnai F20. It successfully finds the optimal defocus and astigmatism with time and accuracy, compared to a human operator.
Keywords:Automatic focusing  Autofocus  Astigmatism  Scanning transmission electron microscopy  HAADF-STEM  Variance  Nelder-Mead simplex method  Derivative-free optimization
本文献已被 ScienceDirect PubMed 等数据库收录!