应用U形边框黑体光阑的三点辐射定标校正方法及其分析
引用本文:修金利,金伟其,刘崇亮,王霞. 应用U形边框黑体光阑的三点辐射定标校正方法及其分析[J]. 红外与激光工程, 2013, 42(9): 2313-2318
作者姓名:修金利  金伟其  刘崇亮  王霞
作者单位:1.北京理工大学 光电学院 “光电成像技术与系统”教育部重点实验室,北京 100081
摘    要:红外焦平面探测器阵列由于探测器工艺、环境冲击和长时间工作等因素将引起探测器响应的漂移,很大程度影响了热成像系统的成像质量。对于红外测温热像仪来说,会大大降低其出厂定标的准确性。针对红外辐射定标,考虑到探测器响应的非线性,在前期搭建的基于U形边框黑体视场光阑的红外成像系统基础上,研究了基于U形边框黑体光阑的三点定标修正方法,并与两点辐射定标方法进行了比较。实际定标测试实验结果表明:在25~65 ℃范围内,三点定标修正后的最大绝对误差和平均误差分别为0.126 6 K和-0.048 8 K,较原始定标的结果有明显的精度提升,说明三点定标修正方法算法有效,但三点定标修正与两点定标修正的结果相差不大。因此,一般情况下两点辐射定标修正方法足以适应辐射定标应用。

关 键 词:红外辐射定标   非线性响应   多项式拟合   三点定标   两点定标
收稿时间:2013-01-07

Three-point infrared radiometric calibration and correction method using U-shaped blackbody
Xiu Jinli , Jin Weiqi , Liu Chongliang , Wang Xia. Three-point infrared radiometric calibration and correction method using U-shaped blackbody[J]. Infrared and Laser Engineering, 2013, 42(9): 2313-2318
Authors:Xiu Jinli    Jin Weiqi    Liu Chongliang    Wang Xia
Affiliation:1.Key Laboratory of Photoelectronic Imaging Technology and System,Ministry of Education of China,School of Optoelectronics,Beijing Institute of Technology,Beijing 100081,China
Abstract:Infrared focal plane array(IRFPA) is an important component of the thermal imaging system. However, due to facts like technical limitations and material defects in production, the environmental influence, and too long time operation, etc., the drift of the IRFPA response during their working is unavoidable. It affects the image quality of the thermal imaging systems and has negative effects on the precision of the thermal imaging equipments. Aiming to solve the problems of traditional radiation calibration and correction methods, considering the nonlinear response of infrared detectors, and taking advantage of the infrared imaging integral platform based on U-shape blackbody, a three-point infrared calibration and correction technology was discussed. Furthermore, the previous tow-point calibration and correction method was used for comparison. The experimental results show that during the temperature range of 25-65 ℃, compared with the original calibration precision, the three-point calibration technology owns quite an effective consequence with the biggest absolute error of 0.126 6 K and average error of -0.048 8 K. However, the method proposed does not have quite abvious difference with the tow-point method, so generally the tow-point calibration and correction can adapt the normal radiometric calibration application.
Keywords:infrared radiometric calibration  nonlinear response  polynomial fitting  three-point calibration  tow-point calibration