1.
Objective

To examine the feasibility of human cardiac MR (CMR) at 14.0 T using high-density radiofrequency (RF) dipole transceiver arrays in conjunction with static and dynamic parallel transmission (pTx).

Materials and methods

RF arrays comprised of self-grounded bow-tie (SGBT) antennas, bow-tie (BT) antennas, or fractionated dipole (FD) antennas were used in this simulation study. Static and dynamic pTx were applied to enhance transmission field (B1+) uniformity and efficiency in the heart of the human voxel model. B1+ distribution and maximum specific absorption rate averaged over 10 g tissue (SAR10g) were examined at 7.0 T and 14.0 T.

Results

At 14.0 T static pTx revealed a minimum B1+ROI efficiency of 0.91 μT/√kW (SGBT), 0.73 μT/√kW (BT), and 0.56 μT/√kW (FD) and maximum SAR10g of 4.24 W/kg, 1.45 W/kg, and 2.04 W/kg. Dynamic pTx with 8 kT points indicate a balance between B1+ROI homogeneity (coefficient of variation < 14%) and efficiency (minimum B1+ROI > 1.11 µT/√kW) at 14.0 T with a maximum SAR10g < 5.25 W/kg.

Discussion

MRI of the human heart at 14.0 T is feasible from an electrodynamic and theoretical standpoint, provided that multi-channel high-density antennas are arranged accordingly. These findings provide a technical foundation for further explorations into CMR at 14.0 T.

  相似文献   

2.
Objective

To measure healthy brain \({T}_{1}\) and \({T}_{2}\) relaxation times at 0.064 T.

Materials and methods

\({T}_{1}\) and \({T}_{2}\) relaxation times were measured in vivo for 10 healthy volunteers using a 0.064 T magnetic resonance imaging (MRI) system and for 10 test samples on both the MRI and a separate 0.064 T nuclear magnetic resonance (NMR) system. In vivo \({T}_{1}\) and \({T}_{2}\) values are reported for white matter (WM), gray matter (GM), and cerebrospinal fluid (CSF) for automatic segmentation regions and manual regions of interest (ROIs).

Results

\({T}_{1}\) sample measurements on the MRI system were within 10% of the NMR measurement for 9 samples, and one sample was within 11%. Eight \({T}_{2}\) sample MRI measurements were within 25% of the NMR measurement, and the two longest \({T}_{2}\) samples had more than 25% variation. Automatic segmentations generally resulted in larger \({T}_{1}\) and \({T}_{2}\) estimates than manual ROIs.

Discussion

\({T}_{1}\) and \({T}_{2}\) times for brain tissue were measured at 0.064 T. Test samples demonstrated accuracy in WM and GM ranges of values but underestimated long \({T}_{2}\) in the CSF range. This work contributes to measuring quantitative MRI properties of the human body at a range of field strengths.

  相似文献   

3.
Magnetic Resonance Materials in Physics, Biology and Medicine - This work investigates electrodynamic constraints, explores RF antenna concepts and examines the transmission fields (B 1 + ) and RF...  相似文献   

4.
Magnetic Resonance Materials in Physics, Biology and Medicine - Given the growing interest in fluorine, it is necessary to develop new multi-tuned RF coils. Therefore, our objective is to design a...  相似文献   

5.
Object

Lower-field MR is reemerging as a viable, potentially cost-effective alternative to high-field MR, thanks to advances in hardware, sequence design, and reconstruction over the past decades. Evaluation of lower field strengths, however, is limited by the availability of lower-field systems on the market and their considerable procurement costs. In this work, we demonstrate a low-cost, temporary alternative to purchasing a dedicated lower-field MR system.

Materials and Methods

By ramping down an existing clinical 3 T MRI system to 0.75 T, proton signals can be acquired using repurposed 13C transmit/receive hardware and the multi-nuclei spectrometer interface. We describe the ramp-down procedure and necessary software and hardware changes to the system.

Results

Apart from presenting system characterization results, we show in vivo examples of cardiac cine imaging, abdominal two- and three-point Dixon-type water/fat separation, water/fat-separated MR Fingerprinting, and point-resolved spectroscopy. In addition, the ramp-down approach allows unique comparisons of, e.g., gradient fidelity of the same MR system operated at different field strengths using the same receive chain, gradient coils, and amplifiers.

Discussion

Ramping down an existing MR system may be seen as a viable alternative for lower-field MR research in groups that already own multi-nuclei hardware and can also serve as a testing platform for custom-made multi-nuclei transmit/receive coils.

  相似文献   

6.
Huang  Zhiwei  Gambarota  Giulio  Xiao  Ying  Wenz  Daniel  Xin  Lijing 《Magma (New York, N.Y.)》2023,36(2):309-315
Purpose

In this study, we aimed to measure the apparent diffusion coefficients (ADCs) of major phosphorous metabolites in the human calf muscle at 7 T with a diffusion-weighted (DW)-STEAM sequence.

Methods

A DW-STEAM sequence with bipolar gradients was implemented at 7 T, and DW MR spectra were acquired in three orthogonal directions in the human calf muscle of six healthy volunteers (TE/TM/TR = 15 ms/750 ms/5 s) at three b-values (0, 800, and 1200 s/mm2). Frequency and phase alignments were applied prior to spectral averaging. Averaged DW MR spectra were analyzed with LCModel, and ADCs of 31P metabolites were estimated.

Results

Four metabolites (phosphocreatine (PCr), adenosine triphosphate (ATP), inorganic phosphate (Pi) and glycerol phosphorylcholine (GPC)) were quantified at all b-values with mean CRLBs below 10%. The ADC values of PCr, ATP, Pi, and GPC were (0.24 ± 0.02, 0.15 ± 0.04, 0.43 ± 0.14, 0.40 ± 0.09) × 10–3 mm2/s, respectively.

Conclusion

The ADCs of four 31P metabolites were successfully measured in the human calf muscle at 7 T, among which those of ATP, Pi and GPC were reported for the first time in humans. This study paves the way to investigate 31P metabolite diffusion properties in health and disease on the clinical MR scanner.

  相似文献   

7.
Introduction

Various research sites are pursuing 14 T MRI systems. However, both local SAR and RF transmit field inhomogeneity will increase. The aim of this simulation study is to investigate the trade-offs between peak local SAR and flip angle uniformity for five transmit coil array designs at 14 T in comparison to 7 T.

Methods

Investigated coil array designs are: 8 dipole antennas (8D), 16 dipole antennas (16D), 8 loop coils (8D), 16 loop coils (16L), 8 dipoles/8 loop coils (8D8L) and for reference 8 dipoles at 7 T. Both RF shimming and kT-points were investigated by plotting L-curves of peak SAR levels vs flip angle homogeneity.

Results

For RF shimming, the 16L array performs best. For kT-points, superior flip angle homogeneity is achieved at the expense of more power deposition, and the dipole arrays outperform the loop coil arrays.

Discussion and conclusion

For most arrays and regular imaging, the constraint on head SAR is reached before constraints on peak local SAR are violated. Furthermore, the different drive vectors in kT-points alleviate strong peaks in local SAR. Flip angle inhomogeneity can be alleviated by kT-points at the expense of larger power deposition. For kT-points, the dipole arrays seem to outperform loop coil arrays.

  相似文献   

8.
Magnetic Resonance Materials in Physics, Biology and Medicine - Speech production MRI benefits from lower magnetic fields due to reduced off-resonance effects at air-tissue interfaces and from the...  相似文献   

9.
Magnetic Resonance Materials in Physics, Biology and Medicine - High resolution MRI of the intracranial vessel wall provides important insights in the assessment of intracranial vascular disease....  相似文献   

10.
Objective

Dysphagia or difficulty in swallowing is a potentially hazardous clinical problem that needs regular monitoring. Real-time 2D MRI of swallowing is a promising radiation-free alternative to the current clinical standard: videofluoroscopy. However, aspiration may be missed if it occurs outside this single imaged slice. We therefore aimed to image swallowing in 3D real time at 12 frames per second (fps).

Materials and methods

At 3 T, three 3D real-time MRI acquisition approaches were compared to the 2D acquisition: an aligned stack-of-stars (SOS), and a rotated SOS with a golden-angle increment and with a tiny golden-angle increment. The optimal 3D acquisition was determined by computer simulations and phantom scans. Subsequently, five healthy volunteers were scanned and swallowing parameters were measured.

Results

Although the rotated SOS approaches resulted in better image quality in simulations, in practice, the aligned SOS performed best due to the limited number of slices. The four swallowing phases could be distinguished in 3D real-time MRI, even though the spatial blurring was stronger than in 2D. The swallowing parameters were similar between 2 and 3D.

Conclusion

At a spatial resolution of 2-by-2-by-6 mm with seven slices, swallowing can be imaged in 3D real time at a frame rate of 12 fps.

  相似文献   

11.
Magnetic Resonance Materials in Physics, Biology and Medicine - An initiative to design and build magnetic resonance imaging (MRI) and spectroscopy (MRS) instruments at 14&nbsp;T and beyond to...  相似文献   

12.
Objective

To determine T1 and T2 relaxation times of healthy pancreas parenchyma at 7 T using a multi-transmit system.

Materials and methods

Twenty-six healthy subjects were scanned with a 7 T MR system using eight parallel transceiver antennas, each with two additional receive loops. A Look-Locker sequence was used to obtain images for T1 determination, while T2 was obtained from spin-echo images and magnetic resonance spectroscopy measurements with different echo times. T1 and T2 times were calculated using a mono-exponential fit of the average magnitude signal from a region of interest in the pancreas and were tested for correlation with age.

Results

The age range of the included subjects was 21–72 years. Average T1 and T2 relaxation times in healthy pancreas were 896 ± 149 ms, and 26.7 ± 5.3 ms, respectively. No correlation with age was found.

Conclusion

T1 and T2 relaxation times of the healthy pancreas were reported for 7 T, which can be used for image acquisition optimization. No significant correlations were found between age and T1 or T2 relaxation times of the pancreas. Considering their low standard deviation and no observable age dependence, these values may be used as a baseline to study potentially pancreatic tissue affected by disease.

  相似文献   

13.
Objective

Amide proton transfer (APT) weighted chemical exchange saturation transfer (CEST) imaging is increasingly used to investigate high-grade, enhancing brain tumours. Non-enhancing glioma is currently less studied, but shows heterogeneous pathophysiology with subtypes having equally poor prognosis as enhancing glioma. Here, we investigate the use of CEST MRI to best differentiate non-enhancing glioma from healthy tissue and image tumour heterogeneity.

Materials & Methods

A 3D pulsed CEST sequence was applied at 3 Tesla with whole tumour coverage and 31 off-resonance frequencies (+6 to -6 ppm) in 18 patients with non-enhancing glioma. Magnetisation transfer ratio asymmetry (MTRasym) and Lorentzian difference (LD) maps at 3.5 ppm were compared for differentiation of tumour versus normal appearing white matter. Heterogeneity was mapped by calculating volume percentages of the tumour showing hyperintense APT-weighted signal.

Results

LDamide gave greater effect sizes than MTRasym to differentiate non-enhancing glioma from normal appearing white matter. On average, 17.9 % ± 13.3 % (min–max: 2.4 %–54.5 %) of the tumour volume showed hyperintense LDamide in non-enhancing glioma.

Conclusion

This works illustrates the need for whole tumour coverage to investigate heterogeneity in increased APT-weighted CEST signal in non-enhancing glioma. Future work should investigate whether targeting hyperintense LDamide regions for biopsies improves diagnosis of non-enhancing glioma.

  相似文献   

14.
Objective To explore the possibilities of proton spectroscopic imaging (1H-MRSI) of the human brain at 7 Tesla with adiabatic refocusing pulses. Materials and methods A combination of conventional slice selective excitation and two pairs of slice selective adiabatic refocusing pulses (semi-LASER) results in the formation of an echo from a localized volume. Depending on the used radio frequency (rf) coil efficiency and available rf power, the duration of the adiabatic full passage pulses (AFPs) is adapted to enable echo times down to 50 ms (head coil) or 30 ms (local surface coil). Results An AFP duration of 5 ms with a corresponding bandwidth of 5.1 kHz resulted in a chemical shift displacement error of 23% over 3.8 ppm at 7T. Using a local surface coil and an echo time down to 30 ms, we detected not only the three main metabolites (NAA, Cr and Cho), but also coupled signals from myo-inositol and glutamate/glutamine in spectra from 0.14 cc voxels with linewidths down to 10 Hz in 10 min measurement time. Conclusions The semi-LASER pulse sequence enables 1H-MRSI of the human brain at 7T for larger parts of the brain as well as small localized areas with both a high spectral and spatial resolution. Part of this work has been presented at the ISMRM-ESMRMB joint meeting in 2007 in Berlin, Germany, abstract number 43.  相似文献   

15.

Objective

Zero echo time (ZTE) and ultrashort echo time (UTE) pulse sequences for MRI offer unique advantages of being able to detect signal from rapidly decaying short-T2 tissue components. In this paper, we applied 3D ZTE and UTE pulse sequences at 7T to assess differences between these methods.

Materials and methods

We matched the ZTE and UTE pulse sequences closely in terms of readout trajectories and image contrast. Our ZTE used the water- and fat-suppressed solid-state proton projection imaging method to fill the center of k-space. Images from healthy volunteers obtained at 7T were compared qualitatively, as well as with SNR and CNR measurements for various ultrashort, short, and long-T2 tissues.

Results

We measured nearly identical contrast-to-noise and signal-to-noise ratios (CNR/SNR) in similar scan times between the two approaches for ultrashort, short, and long-T2 components in the brain, knee and ankle. In our protocol, we observed gradient fidelity artifacts in UTE, and our chosen flip angle and readout also resulted in shading artifacts in ZTE due to inadvertent spatial selectivity. These can be corrected by advanced reconstruction methods or with different chosen protocol parameters.

Conclusion

The applied ZTE and UTE pulse sequences achieved similar contrast and SNR efficiency for volumetric imaging of ultrashort-T2 components. Key differences include that ZTE is limited to volumetric imaging, but has substantially reduced acoustic noise levels during the scan. Meanwhile, UTE has higher acoustic noise levels and greater sensitivity to gradient fidelity, but offers more flexibility in image contrast and volume selection.
  相似文献   

16.
Introduction

MRI of excised hearts at ultra-high field strengths (\({\mathrm{B}}_{0}\)≥7 T) can provide high-resolution, high-fidelity ground truth data for biomedical studies, imaging science, and artificial intelligence. In this study, we demonstrate the capabilities of a custom-built, multiple-element transceiver array customized for high-resolution imaging of excised hearts.

Method

A dedicated 16-element transceiver loop array was implemented for operation in parallel transmit (pTx) mode (8Tx/16Rx) of a clinical whole-body 7 T MRI system. The initial adjustment of the array was performed using full-wave 3D-electromagnetic simulation with subsequent final fine-tuning on the bench.

Results

We report the results of testing the implemented array in tissue-mimicking liquid phantoms and excised porcine hearts. The array demonstrated high efficiency of parallel transmits characteristics enabling efficient pTX-based B1+-shimming.

Conclusion

The receive sensitivity and parallel imaging capability of the dedicated coil were superior to that of a commercial 1Tx/32Rx head coil in both SNR and T2*-mapping. The array was successfully tested to acquire ultra-high-resolution (0.1 × 0.1 × 0.8 mm voxel) images of post-infarction scar tissue. High-resolution (isotropic 1.6 mm3 voxel) diffusion tensor imaging-based tractography provided high-resolution information about normal myocardial fiber orientation.

  相似文献   

17.
T 1-weighted images of the human brain obtained with the MDEFT sequence at 8 T are presented. These images are characterized by an excellent contrast and good signal to noise ratio. Importantly, results were obtained with adiabatic spin inversion and demonstrate that such pulses can be used event in the ultra high frequency (>300 MHz) range. It is thus possible to obtain high quality results at this field strength without violating SAR guidelines.  相似文献   

18.

Object  

To further improve the quality and robustness of the point-spread function (PSF) mapping method for fully automatic and accurate correction of geometric distortions in EPI at ultra high field such as 7 Tesla with high fidelity.  相似文献   

19.
Magnetic Resonance Materials in Physics, Biology and Medicine - Our goal is to design and validate a simple apparatus for the safety assessments of magnetically induced torques by four active...  相似文献   

20.

Object

To design and evaluate a fully shielded, ??/4 stripline resonator as a receive-only surface coil for preclinical MRI at 4.7 T.

Materials and methods

A 20 mm diameter stripline surface coil was fabricated from double-sided Duroid 5880 PCB material and was directly coupled to the input of a MOSFET preamplifier, without requiring a matching network. The new coil was compared with a conventional 20 mm, wire loop, receive-only surface coil in imaging experiments with a separate transmit-only saddle coil.

Results

The stripline surface coil exhibits a loaded Q-factor of 132 at 200 MHz, compared to 138 for a conventional wire loop coil and its resonant frequency drops by 0.2 MHz under loading, rather than 0.5 MHz for the wire loop. The stripline coil displays a more symmetrical B 1 map compared to the wire loop, but the SNR falls off more rapidly with depth so it is 30% poorer 8 mm from the coil plane. It should be possible, however, to reduce this difference by using a thicker dielectric in future versions of the stripline coil.

Conclusion

Compared to a conventional surface coil, the stripline coil is easy to manufacture, requires shorter set-up times and shows reduced dielectric interaction with conductive samples.  相似文献   