The Monarch Chemistry system, a centrifugal analyser incorporating
sophisticated robotics for analytical rotor transfer and flexible
software for workload scheduling, has been evaluated. The optical
system is capable of monitoring absorbance, fluorescence and light
scattering reactions. In addition, an ion selective electrode unit may
be incorporated for the measurement of sodium, potassium and
chloride.The precision, accuracy, linearity, calibration stability and
carry-over were investigated for 19 routine chemistries. The
within-batch and between-day precision data were good in the
majority of cases; some chemistries demonstrated poor performance
at low analyte concentrations. Method comparison studies showed
good agreement, with small discrepancies being due to different
calibration material and methodological differences. Major discrepancies
were found with CK and LD; linearity studies were
good in all cases, except calcium. No significant sample or reagent
carry-over was found.Assessment of throughput for a variety of test profiles varied
between 300 and 605 tests per hour.The instrument was easy to operate, very flexible and capable of
handling a large and varied workload. 相似文献
6.
7.
8.
9.
10.
11.
12.
13.
I Tsuda M Hino T Takubo T Katagami H Kubota S Kawai N Tatsumi 《Journal of Automated Methods and Management in Chemistry》1999,21(4):127-133
The newly developed XE-2100 haematology analyser can provide complete blood counts, leukocyte differentials, perform reticulocyte analysis, and obtain quantitative data on nucleated red blood cells (NRBCs). In this study, we evaluated the basic performance of this instrument using routinely obtained blood specimens treated with ethylenediaminetetraacetic acid-2k. Reproducibility, carryover, stability during storage at 4 degrees C and room temperature, and accuracy were evaluated. In this evaluation, reproducibility was good and little carryover was found. Accurate measurements were possible for up to 48h of storage. A good correlation between findings with the XE-2100 and SE-9000 haematology analysers was found for complete blood count on 210 samples tested. The leukocyte differential obtained with the XE-2100 correlated well with eye counts and with the results obtained with the SE-9000 automated haematology analyser, with r values over 0.9 for the percentages of neutrophils, lymphocytes and eosinophils. The precision and accuracy of VRBC and reticulocyte counts by the XE-2100 were satisfactory. We used the XE-2100 to obtain differential counts for bone marrow aspirates, and good correlations with manual differentials were obtained for total nucleated cell count, percentage of myeloid cells and percentage of erythroid cells. The performance of the XE-2100 was excellent, and this instrument should be able to provide reliable data to clinical laboratories. 相似文献
14.
Taweesook Kanluan Srisanit Intaramanee Surapon Tangvarasittichai Orathai Tangvarasittichai Lerson Suwanton Sudarat Manochiopinij Soontharee Tantrarongroj Pimpan Leangphibul 《Journal of Automated Methods and Management in Chemistry》1991,13(3):97-100
The random access analyser COBAS MIRA S (Roche Diagnostics) was evaluated for two months. The instrument is a computer-controlled discrete analyser which can be run in a combination profile and single test mode. This instrument has
special features, including an automatic cuvette segment changer, a reagent rack cooling system, an external keyboard and monitor, as well as a bar-code facility for the entry of test parameters, worklists and sample identification numbers. Study of within-run and
between-run precision gave values of % CV 0.54-3.37 and 0.61-3.65, respectively, for a variety of assays. Linearity testing to the upper limit of each test was also studied and were found to cover the necessary pathological range. Within the two-month period, no major problems were encountered. The instrument required minimum operator attention during operation. Correlation studies with the Hitachi 705 using six clinical chemistry tests (glucose,
cholesterol, triglyceride, ALP, AST, ALT) gave correlation coefficients ranging from 0.95-0.99 and slopes of 0.91-1.17. 相似文献