1.
This study aims at proposing two TL dosemeters: one for the whole body and another for the extremities, for beta and gamma fields. Selected sensible material consists of 5 mg x cm(-2) LiF:Mg,Cu,P film (GR-200F) manufactured in China. Calibration was carried out according to ISO 4037-3, in terms of Hp(0.07), and dosimetric performance was analysed on the basis of IEC-1066 and ISO-12794 Standards. Experiments showed a satisfactory sensitivity of the proposed dosemeters for detecting beta radiation at protection levels and a very good energy response; thus, highly recommending their use for weakly penetrating radiation measurements. However, the homogeneity and the reproducibility of GR-200F are not found to be as reliable as in standard materials.  相似文献   

2.
3.
Differences and similarities between LiF-based LiF:Mg,Ti and LiF:Mg,Cu,P are discussed, with respect to their dosimetric properties--sensitivity, non-linearity of dose response and heavy charged particle efficiency, as related to the concentration and the individual role of the Mg, Ti, Cu and P dopants. To study further the role of these dopants, the properties of some new, 'hybrid' phosphors: LiF:Mg,Cu,Ti and LiF:Mg,P, specially developed for this purpose, are also discussed. In the glow curve of LiF:Mg,Cu,P with a low concentration of Mg a new peak was found, which appears to be an analogue of peak 4 in LiF:Mg,Ti, Magnesium apparently controls most of the dosimetric properties of LiF-based phosphors. For instance, charged-particle efficiency appears to be anti-correlated with the concentration of Mg, being much less dependent on the content of other dopants. On the other hand, some properties of LiF-based systems seem to be correlated with changes in the emission spectra. It is suggested that Ti hampers the acceptance of any increased amount of Mg into more traps in LiF:MgTi. The absence of Ti, not the presence of P or Cu, is therefore a key to the high sensitivity of LiF:MgCuP.  相似文献   

4.
The ENEA Radiation Protection Institute has been operating the only neutron personal dosimetry service in Italy since the 1970s. Since the 1980s the service has been based on PADC (poly allyl diglycol carbonate) for fast neutron dosimetry, while thermal neutron dosimetry has been performed using thermoluminescence (TL) dosemeters. Since the service was started, a number of aspects have undergone evolution. The latest and most important changes are as follows: in 1998 a new PADC material was introduced in routine, since 2001 TL thermal dosimetry has been based on LiF(Mg,Cu,P) [GR-200] and (7)LiF(Mg,Cu,P) [GR-207] detectors and since 2003 a new image analysis reading system for the fast neutron dosemeters has been used. Herein an updated summary of how the service operates and performs today is presented. The approaches to calibration and traceability to estimate the quantity of H(p)(10) are mentioned. Results obtained at the performance test of dosimetric services in the EU member states and Switzerland sponsored by the European Commission and organised by Eurados in 1999 are reported. Last but not least, quality assurance (QA) procedures introduced in the routine operation to track the whole process of dose evaluation (i.e. plastic QA, acceptance test, test etching bath reproducibility and 'dummy customer' (blind test) for each issuing monitoring period) are presented and discussed.  相似文献   

5.
Evaluation of a new extremity dosemeter is presented. The dosemeter is a passive device that is easy to wear and features a permanent individual numerical ID with barcode, a watertight case, an automatic TLD reader and database management software. Two dosemeters were studied: the first consists of a 100 mg x cm(-2) 7LiF:Mg,Ti (TLD-700) chip and a 42 mg x cm(-2) cap, the other consists of a 7 mg x cm(-2) layer of 7LiF:Mg,Cu,P (TLD-700H) powder and a 5 mg x cm(-2) cap. Sensitivity, repeatability, lower limit detection, angular responses and energy responses for these dosemeters are studied and presented. The dose calculation algorithm is developed and its dosimetric performance accuracy is compared with the standard ANSI N13.32-1995, Performance Testing of Extremity Dosemeters.  相似文献   

6.
LiF:Mg,Cu,P is starting to replace LiF:Mg,Ti in a variety of personnel dosimetry applications. LiF:Mg,Cu,P has superior characteristics as compared to LiF:Mg,Ti including, higher sensitivity, improved energy response for photons, lack of supralinearity and insignificant fading. The use of LiF:Mg,Cu,P in large scale dosimetry programs is of particular interest due to the extreme sensitivity of this material to the maximum readout temperature, and the variety of different dosimetry aspects and details that must be considered for a successful implementation in routine dosimetry. Here we discuss and explain the various aspects of large scale LiF:Mg,Cu,P based dosimetry programs including the properties of the TL material, new generation of TLD readers, calibration methodologies, a new generation of dose calculation algorithms based on the use of artificial neural networks and the overall uncertainty of the dose measurement. The United States Navy (USN) will be the first US dosimetry processor who will use this new material for routine applications. Until June 2002, the Navy used two types of thermoluminescent materials for personnel dosimetry, CaF2:Mn and LiF:Mg,Ti. A program to upgrade the system and to implement LiF:Mg,Cu,P, started in the mid 1990s and was recently concluded. In 2002, the new system replaced the LiF:Mg,Ti and is scheduled to start replacing the CaF2:Mn system in 2006. A pilot study to determine the dosimetric performance of the new LiF:Mg,Cu,P based dosimetry system was recently completed, and the results show the new system to be as good or better than the current system in all areas tested. As a result, LiF:Mg,Cu,P is scheduled to become the primary personnel dosimeter for the entire US Navy in 2006.  相似文献   

7.
The reproducibility of measurements performed with GR-100 (LiF:Mg,Ti) from the Solid Dosimetric Detector and Method Laboratory (DML) China, GR-107 (7LiF:Mg,Ti, DML), TLD-700H (7LiF:Mg.Cu,P, Harshaw) and Al2O3:Mg,Y (Hungary) in photon and mixed photon-neutron fields was investigated. Mixed-field irradiations were performed in a thermal neutron field generated at a nuclear reactor. GR-100 sensitivity decreased after mixed-field irradiations, while no significant change was found for the other materials. Using GR-100 for the dosimetry of mixed and high-intensity fields requires careful procedures.  相似文献   

8.
The effects of UV-induced bleaching of deep traps on Harshaw thermoluminescent (TL) LiF:Mg,Cu,P and LiF:Mg,Ti materials were investigated. During a normal heating cycle, LiF:Mg,Cu,P is limited to a maximum temperature of 240 °C. LiF:Mg,Ti can be read to higher temperatures; however, encapsulation in polytetrafluoroethylene limits the maximum readout temperature to 300 °C. Generally, for both materials, these respective temperatures are sufficient for emptying traps corresponding to the main dosemetric peaks. However, when the dosemeters are subjected to a high dose level, such as 1 Gy (much higher than individual monitoring dose levels), higher temperature traps are filled that cannot be emptied without exceeding the above-mentioned maximum temperatures. These high temperature traps tend to be unstable during normal readout and can significantly increase the residual TL signal. The purpose of this study was to investigate the applicability of a UV-induced bleaching technique for emptying higher temperature traps following high-dose applications. In addition, in the case of LiF:Mg,Cu,P, where the maximum readout temperature is significantly lower, we investigated the possibility of reducing the residual signal using the application of repeated readout cycles. The optical bleaching approach was found to be effective in the case of LiF:Mg,Ti; however, for LiF:Mg,Cu,P, no reduction in the residual signal was observed. For this latter material, the application of repeatable readout cycles is very effective and residual signals equivalent to dose levels as low as 0.01 mGy were observed following an initial dose of 5 Gy. To the best of our knowledge, this work is the first attempt to apply an 'optical annealing' technique to the Harshaw thermoluminescent dosemeter (TLD) materials.  相似文献   

9.
In this paper, the results aimed at assessing the performance of two varieties of LiF detectors (LiF:Mg,Ti and LiF:Mg,Cu,P) in photon fields relatively to reproducibility, detection threshold and angular dependence as defined in the ISO 12794 standard are presented. The fading properties and the limit of detection were also investigated for both materials. The results suggest that both LiF varieties are well suited for extremity monitoring. However, better fading properties of LiF:Mg,Cu,P when compared with LiF:Mg,Ti, combined with previous results relatively to energy dependence suggests that LiF:Mg,Cu,P dosemeters are better suited for extremity monitoring.  相似文献   

10.
In personnel monitoring services, it is important to omit the high-temperature annealing process so that large numbers of TL detectors can be produced economically. There are two efficient ways of reducing the residual signal of LiF:Mg,Cu,P. One is by increasing the maximum readout temperature and the other is by improving the preparation procedure (increasing the Cu concentration and the sintering temperature) but both reduce the TL sensitivity. In personal dosimetry the real dosimetric signals are separated from the residual signals by computerised analysis of glow curves. The adverse influence of the high residual signals of LiF:Mg.Cu.P TL material has been effectively eliminated and the sensitivity remains stable. A good dosimetric result using only reader measurement without pre-irradiation oven annealing is attained in a dose range of 50-80,000 microGy.  相似文献   

11.
LiF:Mg,Cu,P 'pin worms': miniature detectors for brachytherapy dosimetry   总被引:5,自引:0,他引:5  
Dose measurements in brachytherapy 192Ir implants are often difficult due to large dose gradients and complex photon spectra. Therefore, tissue-equivalent detectors with a high spatial resolution, such as the highly promising LiF:Mg,Cu,P thermoluminescent detectors (TLDs) are required. It was the aim of the present work to ascertain if miniature LiF:Mg,Cu,P TLDs can effectively measure the dose distribution around 192Ir implants. 'Pin worm' TLDs (type MCP, diameter 0.6 mm, length 2 mm) were compared with GR-200R (SSDL, Beijing) rods cut in half. The TLDs were tested for reproducibility and energy dependence using high dose rate (HDR) and low dose rate (LDR) brachytherapy units. 192Ir measurements were performed in a tissue equivalent phantom accommodating hollow needles and catheters routinely used in brachytherapy. Pin worms had an average reproducibility of less than +/-2% (1 SD) and a detection limit of less than 10 microGy. The small dimensions of the pin worms allowed their placement within brachytherapy needles and catheters. The measured relative dose distribution was in good agreement with the predictions of a computerised treatment planning system (ADAC Pinnacle); however, limitations in the TLD energy correction did not allow for absolute dose comparison.  相似文献   

12.
The ENEA photon dosemeter, introduced in 1995, consisting of two differently filtrated LiF(Mg,Cu,P) detectors, has been modified recently. The ABS (acrylonitrile butadiene styrene) plastic support has been replaced by a new aluminium card supporting the same two detectors (LiF(Mg,Cu,P) GR200). The new card, fully developed at the ENEA-Radiation Protection Institute (which is going to be patented), can now be processed through a Harshaw Model 6600 Automated TLD Reader, a hot gas reader. This paper reports the results of the individual calibration of approximately 60,000 LiF(Mg,Cu,P) GR200 detectors inserted on the new aluminium cards. Before the implementation in routine of the new cards, the reader has been characterised. Steps and tests to be made to use the card in routine (i.e. reader stability, linearity, reproducibility, etc.) are reported. The whole dosimetric system now combines the very good performances of the Harshaw Model 6600 reader and that of LiF(Mg,Cu,P) thermoluminescent material.  相似文献   

13.
14.
The response of thermoluminescence dosemeters (TLDs) to light, in various conditions, has been studied. TLD cards containing both conventional lithium fluoride (LiF:Mg,Ti) and the high-sensitivity material LiF:Mg,Cu,P were available, so permitting a comparison between the two types. Also available for the tests were Harshaw(TM) extremity EXT-RAD (LiF:Mg,Cu,P) dosemeters. The LiF:Mg,Ti body TLD cards and the EXT-RAD extremity dosemeters both showed some response to fluorescent light, while the LiF:Mg,Cu,P cards showed no significant response. It is therefore concluded that LiF:Mg,Cu,P body cards need no special precautions to protect them from the effects of light. For LiF:Mg,Ti cards and extremity dosemeters, effects are small, but steps to avoid excessive light exposure should be considered.  相似文献   

15.
LiF-based thermoluminescence (TL) materials have been widely used for radiation dosimetry due to their attractive features. LiF:Mg,Cu,P is one of the most sensitive tissue-equivalent TL materials, approximately 40 times more sensitive than LiF:Mg,Ti (TLD-100), but it has two main drawbacks: a thermal loss of the TL sensitivity when annealed at temperatures >240 degrees C, and a relatively high-residual signal. Recently, LiF:Mg,Cu,Na,Si TL material was developed to overcome these drawbacks at the Korea Atomic Energy Research Institute, but it provided only marginal improvements in reducing the residual signal. The newly developed LiF:Mg,Cu,Si TL material has a significantly lower residual signal and a better stability to thermal treatments. In this article, the preparation method and some dosimetric properties (sensitivity and residual signal) of the new LiF:Mg,Cu,Si TL material are presented. At the end of the preparation procedures, a dual-step annealing method is introduced and this has proved as a very efficient method to reduce the high-temperature peak and is the cause of residual signal. Therefore, the high-temperature peak in the glow curve was significantly reduced. The sensitivity is approximately 20 times higher than that of TLD-100 and the residual signal was estimated to be approximately 0.04%.  相似文献   

16.
Sintered LiF:Mg,Cu,Na,Si thermoluminescence (TL) pellets have been developed for application in radiation dosimetry. LiF:M,Cu,Na,Si TL pellets were made from TL powders using a sintering process, that is, pressing and heat treatment. These pellets have a diameter of 4.5 mm, and a thickness of 0.8 mm are blue in colour and have a mass of 28 mg each. After 400 pellets had been produced they were irradiated with 137Cs gamma radiation and samples having a sensitivity within a +/-5% standard deviation were selected for experimental use. In the present study, the physical and dosimetric properties of LiF:Mg,Cu,Na,Si TL pellets were investigated for their emission spectrum, dose response, energy response and fading characteristics. Photon irradiation for the experiments was carried out using X ray beams and a 137Cs gamma source at the Korea Atomic Energy Research Institute (KAERI). The average energies and the dose were in the range of 20-662 keV and 10(-6) - 10(2) Gy respectively. The glow curves were measured with a manual type thermoluminescence dosimetry reader (system 310, Teledyne) at a constant nitrogen flux and a linear heating rate. For a constant heating rate of 5 degrees C.s(-1). the main dosimetric peak of the glow curve appeared at 234 degrees C, its activation energy was 2.34 eV and the frequency factor was 1.00 x 10(23). The TL emission spectrum appeared at the blue region centred at 410 nm. A linearity of photon dose response was maintained up to 100 Gy. The photon energy responses relative to the 137Cs response were within +/-20% in the overall photon energy region. No fading of the TL sensitivity of the pellets stored at room temperature was found over the course of a year. Therefore LiF:Mg,Cu,Na,Si TL pellets can be used for personal dosimetry, but more research is needed to improve the characteristics for repeated use.  相似文献   

17.
Recently, two new types of 'tissue equivalent' thermoluminescent detectors (TLDs) have aroused attention: LiF:Mg,Cu,Na,Si and Li2B4O7:Cu,Ag,P. In this work the characteristics of both detectors were compared with the characteristics of the well-known type LiF:Mg,Ti detector, TLD-100. The following properties were investigated: the glow curve structures, relative sensitivity, batch homogeneity and uniformity, detection threshold, reproducibility of the response, linearity in the wide dose range and fading. Also, the energy dependence for medium and low energy X rays was determined in the range of mean energies between 33 and 116 keV. The results confirmed 'tissue equivalency' of both new types in the investigated range of photon energies. LiF:Mg,Cu,Na,Si detector has very high sensitivity (approximately 75 times higher than that of TLD-100) and is convenient for use in a very low range of doses. Li2B4O7:Cu,Ag,P detector shows some improvements in comparison with the previously prepared types of lithium borate. The most important is the five times higher sensitivity than that of TLD-100. This detector is also very promising, especially in medical dosimetry.  相似文献   

18.
This work presents an extensive study on Monte Carlo radiation transport simulation and thermoluminescent (TL) dosimetry for characterising mixed radiation fields (neutrons and photons) occurring in nuclear reactors. The feasibility of these methods is investigated for radiation fields at various locations of the Portuguese Research Reactor (RPI). The performance of the approaches developed in this work is compared with dosimetric techniques already existing at RPI. The Monte Carlo MCNP-4C code was used for a detailed modelling of the reactor core, the fast neutron beam and the thermal column of RPI. Simulations using these models allow to reproduce the energy and spatial distributions of the neutron field very well (agreement better than 80%). In the case of the photon field, the agreement improves with decreasing intensity of the component related to fission and activation products. (7)LiF:Mg,Ti, (7)LiF:Mg,Cu,P and Al(2)O(3):Mg,Y TL detectors (TLDs) with low neutron sensitivity are able to determine photon dose and dose profiles with high spatial resolution. On the other hand, (nat)LiF:Mg,Ti TLDs with increased neutron sensitivity show a remarkable loss of sensitivity and a high supralinearity in high-intensity fields hampering their application at nuclear reactors.  相似文献   

19.
A method of measurement of radon concentration in air was developed, based on high-sensitivity LiF:Mg,Cu,P (MCP-N, TLD Poland) thermoluminescent detectors installed in charcoal canisters. The canisters were exposed typically for 72 h in a calibration chamber with a radon concentration ranging from 100 Bq x m(-3) to 87 kBq x m(-3). It was found that in these conditions the signal registered by the TL detectors was proportional to the 222Rn concentration and the lowest limit of detection (LLD) was at a level of 100 Bq x m(-3). The proposed method can be used in large-scale, multi-site surveys aimed at screening for high levels of indoor radon concentration or for measuring ground radon exhalation rates.  相似文献   

20.
Low-energy alpha particle and proton heavy charged particle (HCP) relative thermoluminescence (TL) efficiencies are calculated for the major dosimetric glow peak in LiF:Mg,Cu,P (MCP-N) in the framework of track structure theory (TST). The calculations employ previously published TRIPOS-E Monte Carlo track segment values of the radial dose in condensed phase LiF calculated at the Instituto National de Investigaciones Nucleares (Mexico) and experimentally measured normalised (60)Co gamma-induced TL dose-response functions, f(D), carried out at the Institute of Nuclear Physics (Poland). The motivation for the calculations is to test the validity of TST in a TL system in which f(D) is not supralinear (f(D) >1) and is not significantly dependent on photon energy contrary to the behaviour of the dose-response of composite peak 5 in the glow curve of LiF:Mg,Ti (TLD-100). The calculated HCP relative efficiencies in LiF:MCP-N are 23-87% lower than the experimentally measured values, indicating a weakness in the major premise of TST which exclusively relates HCP effects to the radiation action of the secondary electrons liberated by the HCP slowing down. However, an analysis of the uncertainties involved in the TST calculations and experiments (i.e. experimental measurement of f(D) at high levels of dose, sample light self-absorption and accuracy in the estimation of D(r), especially towards the end of the HCP track) indicate that these may be too large to enable a definite conclusion. More accurate estimation of sample light self-absorption, improved measurements of f(D) and full-track Monte Carlo calculations of D(r) incorporating improvements of the low-energy electron transport are indicated in order to reduce uncertainties and enable a final conclusion.  相似文献   