Wang  Jie  Gao  Da-Qing  Shen  Wan-Zeng  Yan  Hong-Bin 《核技术(英文版)》2022,33(10):1-12

The Shanghai advanced proton therapy facility is a proton cancer treatment device designed and built by the Shanghai Institute of applied physics at the Chinese academy of sciences. The accelerator part comprises a proton linear accelerator (linac) injector and a circular synchrotron. An alternating current current transformer (ACCT) is used for non-intercepting beam current measurement at the drift tube linac exit. According to the beam characteristics, the ACCT is required to complete real-time beam current and pulse width measurements at currents of 3–30 mA, frequencies of 1–10 Hz, and pulse widths of 40–400 μs. In this paper, we report the design and development of an ACCT. The designed ACCT was simulated using CST Microwave Studio, and calibrated using an oscilloscope and signal generator. Variations in the output signal of the ACCT were investigated as a function of ceramic gap size, number of coil turns, and resistance. According to the simulation and experimental results, the optimal number of coil turns was found to be 30. In addition, a low-pass filter was adopted to filter the noise introduced during long-distance signal transmission using a coaxial cable with the length of 20 m. The calibration results show that the corresponding rise time of the ACCT is 800 ns with the sensitivity of 8.2 V/A and a droop of less than 1%, meeting the design requirements.