为进一步提升HLS Ⅱ储存环光源的性能以及为未来实现top-off注入打下基础,研制了新的逐束团流强测量系统。系统硬件部分主要由ADC、FPGA和USB构成,采样精度高、架构简单、成本低。由于BPM和信号的峰值含有束团流强信息,该系统利用峰值采样法测流强。在线测试结果表明,束流的纵向振荡对测量精度有较大影响,在开启纵向反馈抑制住纵向振荡时,该系统测量逐束团流强的精度较高,单束团流强的均方根误差值可达0.002 m A;关闭纵向反馈时,测量精度变差。该系统除了测量逐束团流强外,还实现了纵向工作点的测量,未来还可以测量逐束团寿命等,为合肥光源储存环性能提升,提供更多的束流诊断手段。 相似文献
3.
The four electrodes in the stripline beam position monitor(BPM) for Hefei Light Source(HLS II) storage ring are of axially symmetric type. We have derived a new calibration method of electrode gains for this type stripline BPM. The gain fit error of different data grids was analyzed, and the ±5 mm by ±5 mm grid is the best.The electrode gains of two stripline BPMs(HLS II SR-BD-STLB1 and HLS II SR-BD-STLB2) were obtained based on offline calibrated data. The results show that data after fitting gains are improved, with the electrode gains being between 0.94 and 1.15. 相似文献
4.
一种利用可见光的快速束团横向尺寸检测器(Fast Beam Transverse Profile Monitor,FBPM)正在研制过程中,并用于测量快速束流位置和尺寸,该装置将应用到升级改造完成后的合肥光源。FBPM基于日本滨松公司生产的多阳极光电倍增管(Multi-Anode Photo-Multiplier Tube,MAPMT)R5900U-00-L16,它每个通道拥有的有效感光区域为0.8 mm?16 mm,典型上升时间为0.6 ns,因此拥有足够的空间分辨率和时间分辨率。本文介绍了同步光成像系统,设计了低噪声宽带前置放大器,基于光电倍增管4个相邻通道设计了对数处理算法。离线测试结果表明,透镜焦距的测量误差小于?1 mm,放大器带宽达到420 MHz,4个通道相对参考通道2的增益偏差小于0.2 d B,4通道峰值信号偏差不超过2%。 相似文献