LI Zhaobin XIONG Fei HUANG Guofeng CAO Zheng JIANG Ruiyao FU Shen 《核技术(英文版)》2007,18(3):145-149
Linear accelerators equipped with multileaf collimators (MLC) are becoming more common and are widely used in the intensity-modulated radiotherapy (IMRT). There is an imperative need to ensure the commissioning specification of the linear accelerators for the sake of quality assurance (QA) and quality control (QC). This paper is aimed to investigate the role of AGFA high-energy CR (Computed Radiography) in calibrating dynamic multileaf collimators and evaluating the accuracy of the leaf position. The result shows that AGFA high-energy CR can easily and conveniently be used to calibrate MLC and verify its position. Hence, the application of AGFA high-energy CR is proved to be an accurate and time-saving method for routine MLC QC, especially when MLC calibration adjustments are required. 相似文献