Response of neutron detectors to high-energy mixed radiation fields |
| |
Authors: | Mayer S Forkel-Wirth D Fuerstner M Menzel H G Mueller M J Perrin D Theis C Vincke H |
| |
Affiliation: | Paul Scherrer Institut CH-5232 Villigen PSI, Switzerland. |
| |
Abstract: | Radiation protection around CERN's high-energy accelerators represents a major challenge due to the presence of complex, mixed radiation fields. Behind thick shielding neutrons dominate and their energy ranges from fractions of eV to about 1 GeV. In this work the response of various portable detectors sensitive to neutrons was studied at CERN's High-Energy Reference Field Facility (CERF). The measurements were carried out with conventional rem counters, which usually cover neutron energies up to 20 MeV, the Thermo WENDI-2, which is specified to measure neutrons up to several GeV, and a tissue-equivalent proportional counter. The experimentally determined neutron dose equivalent results were compared with Monte Carlo (MC) simulations. Based on these studies field calibration factors can be determined, which result in a more reliable estimate of H(*)(10) in an unknown, but presumably similar high-energy field around an accelerator than a calibration factor determined in a radiation field of a reference neutron source. |
| |