1.Russian Federal Nuclear Center—Zababakhin All-Russian Scientific Research Institute of Technical Physics,Snezhinsk, Chelyabinsk oblast,Russia
Abstract:
The thickness gauging of nonmagnetic metal coatings applied to nonmagnetic bases with tuning out from base electric conductivity is considered. A measurement technique that enables the tuning out in increased-locality measurements within a wide range of coating thicknesses is described. Results of computational and experimental research are provided. Optimum inspection parameters are analyzed.