首页 | 本学科首页   官方微博 | 高级检索  
     


Relaxation measurements of an MRI system phantom at low magnetic field strengths
Authors:Martin  Michele N  Jordanova  Kalina V  Kos  Anthony B  Russek  Stephen E  Keenan  Kathryn E  Stupic  Karl F
Affiliation:1.U.S. Department of Commerce, National Institute of Standards and Technology, 325 Broadway, Boulder, CO, 80305, USA
;
Abstract:Objective

Temperature controlled T1 and T2 relaxation times are measured on NiCl2 and MnCl2 solutions from the ISMRM/NIST system phantom at low magnetic field strengths of 6.5 mT, 64 mT and 550 mT.

Materials and methods

The T1 and T2 were measured of five samples with increasing concentrations of NiCl2 and five samples with increasing concentrations of MnCl2. All samples were scanned at 6.5 mT, 64 mT and 550 mT, at sample temperatures ranging from 10 °C to 37 °C.

Results

The NiCl2 solutions showed little change in T1 and T2 with magnetic field strength, and both relaxation times decreased with increasing temperature. The MnCl2 solutions showed an increase in T1 and a decrease in T2 with increasing magnetic field strength, and both T1 and T2 increased with increasing temperature.

Discussion

The low field relaxation rates of the NiCl2 and MnCl2 arrays in the ISMRM/NIST system phantom are investigated and compared to results from clinical field strengths of 1.5 T and 3.0 T. The measurements can be used as a benchmark for MRI system functionality and stability, especially when MRI systems are taken out of the radiology suite or laboratory and into less traditional environments.

Keywords:
本文献已被 SpringerLink 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号