首页 | 本学科首页   官方微博 | 高级检索  
     


Radiofrequency antenna concepts for human cardiac MR at 14.0 T
Authors:Nurzed  Bilguun  Kuehne  Andre  Aigner  Christoph Stefan  Schmitter  Sebastian  Niendorf  Thoralf  Eigentler  Thomas Wilhelm
Affiliation:1.Max-Delbrück-Center for Molecular Medicine in the Helmholtz Association (MDC), Berlin Ultrahigh Field Facility (B.U.F.F.), Robert Rössle Strasse 10, 13125, Berlin, Germany
;2.MRI.TOOLS GmbH, Berlin, Germany
;3.Physikalisch-Technische Bundesanstalt (PTB), Braunschweig, Berlin, Germany
;4.Experimental and Clinical Research Center (ECRC), a joint cooperation between the Charité Medical Faculty and the Max-Delbrück-Center for Molecular Medicine in the Helmholtz Association, Berlin, Germany
;5.Chair of Medical Engineering, Technische Universität Berlin, Berlin, Germany
;
Abstract:Objective

To examine the feasibility of human cardiac MR (CMR) at 14.0 T using high-density radiofrequency (RF) dipole transceiver arrays in conjunction with static and dynamic parallel transmission (pTx).

Materials and methods

RF arrays comprised of self-grounded bow-tie (SGBT) antennas, bow-tie (BT) antennas, or fractionated dipole (FD) antennas were used in this simulation study. Static and dynamic pTx were applied to enhance transmission field (B1+) uniformity and efficiency in the heart of the human voxel model. B1+ distribution and maximum specific absorption rate averaged over 10 g tissue (SAR10g) were examined at 7.0 T and 14.0 T.

Results

At 14.0 T static pTx revealed a minimum B1+ROI efficiency of 0.91 μT/√kW (SGBT), 0.73 μT/√kW (BT), and 0.56 μT/√kW (FD) and maximum SAR10g of 4.24 W/kg, 1.45 W/kg, and 2.04 W/kg. Dynamic pTx with 8 kT points indicate a balance between B1+ROI homogeneity (coefficient of variation?<?14%) and efficiency (minimum B1+ROI?>?1.11 µT/√kW) at 14.0 T with a maximum SAR10g?<?5.25 W/kg.

Discussion

MRI of the human heart at 14.0 T is feasible from an electrodynamic and theoretical standpoint, provided that multi-channel high-density antennas are arranged accordingly. These findings provide a technical foundation for further explorations into CMR at 14.0 T.

Keywords:
本文献已被 SpringerLink 等数据库收录!