首页 | 本学科首页   官方微博 | 高级检索  
     


Biosensor-based determination of folic acid in fortified food
Authors:Malin Bostr  m Caselunghe,Johan Lindeberg
Affiliation:

a Biacore AB, Rapsgatan 7, SE-754 50 Uppsala, Sweden

b National Food Administration, Box 622, SE-751 26 Uppsala, Sweden

Abstract:A newly developed method for the quantification of folic acid in fortified food is presented. An immunoaffinity-based optical biosensor was used to determine folic acid concentration levels in milk powder, infant formula and cereal samples. Accuracy of the method (88–101%) was demonstrated with the analysis of five reference samples. A collaborative precision study, where ten participants at four different laboratories analysed a set of ten samples, resulted in repeatability relative standard deviations of 2–8% and reproducibility relative standard deviations of 4–10%.
Keywords:Biosensors   Folic acid   Fortified food   Analyses
本文献已被 ScienceDirect 等数据库收录!