Sulfur-specific detection of impurities in cimetidine drug substance using liquid chromatography coupled to high resolution inductively coupled plasma mass spectrometry and electrospray mass spectrometry |
| |
Authors: | Evans E H Wolff J C Eckers C |
| |
Affiliation: | Plymouth Environmental Research Centre, Department of Environmental Sciences, University of Plymouth, Devon, UK. hevans@plymouth.ac.uk |
| |
Abstract: | The use of liquid chromatography coupled to sector field inductively coupled plasma mass spectrometry (SF-ICP-MS) for the specific detection of sulfur-containing compounds is described. In the sulfur-containing drug substance cimetidine, structurally related impurities well below the 0.1% mass fraction level relative to the main drug substance could easily be detected. The structure of most of the impurities was confirmed by electrospray mass spectrometry (ESI-MS), and thus, the complementarity of the two techniques for drug analysis is shown. The limit of detection by SF-ICP-MS for cimetidine in solution was approximately 4-20 ng x g(-1), but it was blank-limited. |
| |