Phase synchronization of an ion ensemble by frequency sweep excitation in Fourier transform ion cyclotron resonance |
| |
Authors: | C D Hanson M E Castro D H Russel |
| |
Abstract: | The Fourier transform ion cyclotron resonance (FT-ICR) signal is produced by the coherent motion of a population of ions. The ability to produce a well-defined ion packet by excitation of an initially random ion ensemble is a major limiting factor of high mass FT-ICR. Ions must be both resonant and in phase with the applied radio frequency excitation field to be accelerated to radii suitable for detection by FT-ICR. Synchronization of the phase angles of an ensemble of ions occurs by off-resonant acceleration during frequency swept excitation. Results from computer-simulated ion trajectories suggest that phase synchronization of the ion packet prior to resonant excitation results in better spatial definition of the ion ensemble. |
| |