首页 | 本学科首页   官方微博 | 高级检索  
     

g 值加速度校准激励脉冲的横向激光干涉测量方法
引用本文:范锦彪,李玺,徐鹏,祖静.高g 值加速度校准激励脉冲的横向激光干涉测量方法[J].红外与激光工程,2015,44(2):497-502.
作者姓名:范锦彪  李玺  徐鹏  祖静
作者单位:1.中北大学电子测试技术重点实验室,山西太原030051;
基金项目:武器装备预研基金(9140C1204050609);山西省青年科学基金
摘    要:为了准确测量高g值加速度传感器校准激励脉冲的幅值和持续时间,研究了基于激光多普勒的加速度脉冲测量方法。在分析激光多普勒原理的基础上,设计了以光栅为合作目标的激光测速仪;针对干涉仪输出的存在零点漂移的多普勒信号,提出了基于多项式拟合和希尔伯特变换的局部均值分解法;利用该方法对实测多普勒信号进行了解算,得到了激励加速度脉冲;分析了由干涉仪及其合作目标引起的加速度脉冲的幅值和脉宽测量不确定度。实测数据表明:该方法在包含因子k=2时幅值测量的扩展不确定度为U=3%,脉宽测量的扩展不确定度为U=4.8%。测量精度符合ISO16063-13(2001)规定的指标。

关 键 词:横向干涉测量法    校准激励脉冲    幅值    脉宽    高g  值加速度传感器
收稿时间:2014-06-11

Measurement of exciting pulse for calibration of high- g accelerometer using lateral laser interferometer
Fan Jinbiao,Li Xi,Xu Peng,Zu Jing.Measurement of exciting pulse for calibration of high- g accelerometer using lateral laser interferometer[J].Infrared and Laser Engineering,2015,44(2):497-502.
Authors:Fan Jinbiao  Li Xi  Xu Peng  Zu Jing
Affiliation:1.Science and Technology on Electronic Test & Measurement Laboratory,North University of China,Taiyuan 030051,China;2.Key Laboratory of Instrumentation Science & Dynamic Measurement,Ministry of Education,North University of China Taiyuan 030051,China;3.College of Computer and Control Engineering,North University of China Taiyuan 030051,China;4.School of Science,North University of China,Taiyuan 030051,China
Abstract:To accurately measure the amplitude and duration of exciting pulse for calibration of a high-g accelerometer, the measurement method of the exciting pulse was studied in this paper. Based on the analysis of the principle of Laser Doppler, a grating laser velocimeter was designed. A local mean decomposition algorithm based on polynomial fitting and Hilbert transform was put forward to process the Doppler signal with zero shift. The acceleration pulse was calculated using the above algorithm and the uncertainties of measurement of the amplitude and duration were analyzed and calculated. The expanded uncertainties verified by experiments are 3% for the amplitude and 4.8% for the duration with a coverage factor k =2. This method is effective for pulse measurement and the measurement precision met the requirements of ISO16063-13(2001).
Keywords:transverse interferometry  exciting pulse  amplitude  duration  high- g accelerometer
本文献已被 CNKI 万方数据 等数据库收录!
点击此处可从《红外与激光工程》浏览原始摘要信息
点击此处可从《红外与激光工程》下载全文
设为首页 | 免责声明 | 关于勤云 | 加入收藏