首页 | 本学科首页   官方微博 | 高级检索  
     


Sectionable microcarrier beads: An improved method for the preparation of cell monolayers for electron microscopy
Authors:M E Hogan  H I Hassouna  K L Klomparens
Abstract:Cross-linked dextran beads provide an excellent surface for tissue-cultured cell monolayers, and can be processed for transmission (TEM) and scanning (SEM) electron microscopy, as well as light microscopy (LM). Cells are grown to confluency on the surface of the microcarriers, where at any point aliquots can be removed and experimentally treated as desired (e.g. immunocytochemistry) providing a representative sample. Sample preparation for TEM follows standard procedures for any cell monolayer, but infiltration times must be at least doubled to allow penetration of the beads. The polymerized blocks can then be sectioned for TEM or LM with no additional steps required. SEM sample preparation involves attaching the fixed bead/cell suspension to a glass coverslip with poly-1-lysine, dehydration, critical point drying, and coating for conductivity. The fixed and dried sample can also be attached directly to the SEM stub as free beads and subsequently gold coated. These beads provide (1) an increased surface area of cells visible per area of thin section, (2) eliminates the careful orientation required for flat substrate methods of embedding, (3) decreases the amount of sample manipulation in the forms of re-embedding and gluing, and (4) decreases the amount of drying artifact seen as cracking in SEM monolayer preparations.
Keywords:Monolayer cells  Microcarrier beads  Transmission electron microscopy  Scanning electron microscopy