首页 | 本学科首页   官方微博 | 高级检索  
     


Microfluidic transendothelial electrical resistance measurement device that enables blood flow and postgrowth experiments
Authors:Vogel Paul A  Halpin Stephen T  Martin R Scott  Spence Dana M
Affiliation:Department of Chemistry, Michigan State University, East Lansing, Michigan 48824, USA.
Abstract:Transendothelial electronic resistance (TEER) measurements are performed across a cell layer immobilized on a microfluidic device that also enables the cell layer to interact with a flowing stream of red blood cells (RBCs). A bipolar pulsed square wave potential is applied across a monolayer of bovine pulmonary artery endothelial cells, and the resulting current response is measured and integrated. The overall impedance of the cell layer provides an indicator of cell layer integrity. After cell seeding on the device, a decrease in TEER signal from 22.3 ± 1.6 μC to 3.5 ± 0.4 μC (corresponding to a resistance of 40.9 ± 2.9 Ω·cm(2) to 259.1 ± 27.4 Ω·cm(2)) was observed after 8 h of cell growth. Intracellular nitric oxide (NO) production by the immobilized endothelial cells that had reached confluence was 34% higher than those cells that had not reached confluence, as indicated by the integrated TEER system. Importantly, this NO production by the confluent endothelium was stimulated by ATP released from RBCs flowing under the endothelial cells. In this construct, the described microfluidic device enables both a TEER-based evaluation of cell layer integrity and molecularly communicated interactions of these cells with a flowing stream of blood components.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号