首页 | 本学科首页   官方微博 | 高级检索  
     


Measurement and estimation of temperature rise in TEM sample during ion milling
Authors:Park Young Min  Ko Dong-Su  Yi Kyung-Woo  Petrov Ivan  Kim Young-Woon
Affiliation:Department of Materials Science and Engineering, Seoul National University, Sillim9-dong Gwanak-gu, Seoul 151-744, Republic of Korea.
Abstract:The actual temperature rise was measured during ion-milling process used in the transmission electron microscopy (TEM) sample preparation. Special probes were fabricated for the measurements, one with shielded, floating thermocouple mounted onto a 3mm grid to compute the thermal load at the sample, and the other, a bare probe with a polymer coating to measure the maximum temperature attained. The temperature measured in the most commonly used ion-milling system reached up to 295 degrees C when the typical milling conditions, 6keV ion-energy and an incident angle of 80 degrees, were used. Based on the temperature profiles that were obtained by the shielded probe, two unknown parameters, the amount of heat deposited by the energetic ions/neutrals to the sample and the thermal conductivities between the materials, were estimated and used to compute the temperature rise in commonly adopted materials. The calculation showed that the temperature of the glass sample reached more than 300 degrees C under typical ion-milling conditions. The calculated value was confirmed with the experimental result of the crystallization of an amorphous Si on the glass under the typical ion-milling condition, which gave the same extent as annealing at 350 degrees C.
Keywords:61  16  Bg  61  80  Jh  81  05  Kf
本文献已被 ScienceDirect PubMed 等数据库收录!