首页 | 本学科首页   官方微博 | 高级检索  
     


Development of the Sealed-Type Triple-Point-of-Argon Cell for Long-Stem SPRT Calibration at KRISS
Authors:I. Yang  C. H. Song  K. H. Kang  Y. -G. Kim  K. S. Gam
Affiliation:(1) Division of Physical Metrology, Korea Research Institute of Standards and Science (KRISS), 1 Doryong-dong, Yuseong-gu, Daejeon, 305-340, South Korea
Abstract:A system was fabricated to realize the triple point of argon for the calibration of long-stem standard platinum resistance thermometers. A cryostat was constructed so that the temperature could be controlled quasi-adiabatically, and the melting was realized using the continuous-heating method. The combined uncertainty of the realization of the triple point of argon for a confidence level of 95% was 0.6 mK.
Keywords:Calibration  Long-stem SPRT  Triple point of argon  Thermometry