首页 | 本学科首页   官方微博 | 高级检索  
     


Essential experimental parameters for quantitative structure analysis using spherical aberration-corrected HAADF-STEM
Authors:Yasutoshi Kotaka
Affiliation:Device & Materials Laboratories, Fujitsu Laboratories Ltd., 10-1 Morinosato-Wakamiya, Atsugi 243-0197, Japan
Abstract:The accuracy of quantitative analysis for Z-contrast images with a spherical aberration (Cs) corrected high-angle annular dark-field (HAADF) scanning transmission electron microscope (STEM) using SrTiO3(0 0 1) was systematically investigated. Atomic column and background intensities were measured accurately from the experimental HAADF-STEM images obtained under exact experimental condition. We examined atomic intensity ratio dependence on experimental conditions such as defocus, convergent semi-angles, specimen thicknesses and digitalized STEM image acquisition system: brightness and contrast. In order to carry out quantitative analysis of Cs-corrected HAADF-STEM, it is essential to determine defocus, to measure specimen thickness and to fix setting of brightness, contrast and probe current. To confirm the validity and accuracy of the experimental results, we compared experimental and HAADF-STEM calculations based on the Bloch wave method.
Keywords:HAADF-STEM  Cs-corrected STEM  Z-contrast
本文献已被 ScienceDirect 等数据库收录!