Data handling in g.p.c. for routine operations |
| |
Authors: | J.P Busnel |
| |
Affiliation: | Laboratoire de Physico-chimie macromoléculaire, E.R.A. 311 - Faculté des Sciences - 72017 Le Mans Cedex, France |
| |
Abstract: | The main points of importance in g.p.c. for routine operation are discussed. Concerning reproductibility, the major problem is the elution volume control, several methods are compared; the influence of number and precision of detector signal measurements are presented. For calibration, working with THF as the eluant, the use of narrow fraction polystyrene standards is a necessary primary step. It furnishes, by classical summations, number and weight average molecular weights in equivalent polystyrenes; these values are already useful in sample characterization. In estimating true molecular weights, the simplest method is use of the Benoit factor and some practical applications are given. A simple method for taking axial dispersion into account is proposed where the spreading function is not Gaussian. |
| |
Keywords: | Chromatographic analysis gel permeation chromatography calibration characterization polystyrene Benoit factor axial dispersion |