首页 | 本学科首页   官方微博 | 高级检索  
     


Performance of a long-wave infrared hyperspectral imager using a Sagnac interferometer and an uncooled microbolometer array
Authors:Lucey Paul G  Horton Keith A  Williams Tim
Affiliation:Hawaii Institute of Geophysics and Planetology, University of Hawaii at Manoa, 1680 East West Road, Honolulu, Hawaii 96822, USA. lucey@higp.hawaii.edu
Abstract:Field and laboratory measurements using an interferometer spectrometer based on the Sagnac interferometer using a microbolometer array detector are presented. Remotely obtained signatures collected with this instrument and with a cryogenic IR spectrometer are compared and shown to closely correspond. Ground-to-ground and air-to-ground image products are presented that demonstrate the image quality of the sensor. Signal-to-noise measurements are presented and compared with a simple parametric performance model that predicts the sensor performance. The performance model is used to predict the performance of this technology when equipped with cooled detectors.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号