首页 | 本学科首页   官方微博 | 高级检索  
     


Trilateral Intercomparison of Photometric Units Maintained at NIST (USA), NPL (UK), and PTB (Germany)
Authors:Yoshi Ohno  Teresa Goodman  Georg Sauter
Affiliation:National Institute of Standards and Technology, Gaithersburg, MD 20899-8442 USA;National Physical Laboratory, Queens Road, Teddington, Middlesex TW11 0LW UK;Physikalisch-Technische Bundesanstalt, Abteilung Optik, Photometrie, Postfach 33 45, 38023 Braunschweig, Germany
Abstract:A trilateral intercomparison of photometric units between NIST (USA), NPL (UK), and PTB (Germany) has been conducted to update the knowledge of the relationship between the photometric units disseminated in the three countries. The luminous intensity unit (cd), the luminous responsivity scale (A/lx), and the luminous flux unit (lm) maintained at each laboratory were compared by circulating transfer standard lamps and photometers. The results showed that the relative luminous intensity values, with respect to the average, measured by NIST, NPL, and PTB were 1.0014, 1.0021, and 0.9966; the relative inverse values of the luminous responsivity (corresponding to illuminance) were 1.0023, 1.0011, and 0.9965; the relative luminous flux values were 0.9994, 1.0034, and 0.9972, respectively. The results agreed within the stated uncertainties of the units maintained at the three laboratories.
Keywords:illuminance  intercomparison  luminous flux  luminous intensity  luminous responsivity  photometer  photometry  units
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号