Optimization of a bolometer detector for ITER based on Pt absorber on SiN membrane
Authors:Meister H,Eich T,Endstrasser N,Giannone L,Kannamüller M,Kling A,Koll J,Trautmann T  ASDEX Upgrade Team,Detemple P,Schmitt S
Affiliation:Max-Planck-Institut für Plasmaphysik, EURATOM Association, Boltzmannstr. 2, D-85748 Garching, Germany. meister@ipp.mpg.de
Abstract:Any plasma diagnostic in ITER must be able to operate at temperatures in excess of 200?°C and neutron loads corresponding to 0.1 dpa over its lifetime. To achieve this aim for the bolometer diagnostic, a miniaturized metal resistor bolometer detector based on Pt absorbers galvanically deposited on SiN membranes is being developed. The first two generations of detectors featured up to 4.5?μm thick absorbers. Results from laboratory tests are presented characterizing the dependence of their calibration constants under thermal loads up to 450?°C. Several detectors have been tested in ASDEX Upgrade providing reliable data but also pointing out the need for further optimization. A laser trimming procedure has been implemented to reduce the mismatch in meander resistances below 1% for one detector and the thermal drifts from this mismatch.