首页 | 本学科首页   官方微博 | 高级检索  
     


A simple technique to measure the movements of the microscope stage along the x and y axes for stereological methods
Authors:S Kaplan  S Canan  H Aslan  B Ünal  & B Sahi&#;n
Affiliation:Departments of Histology-Embryology, Physiology and Anatomy, Medical Faculty, Ondokuz May?s University, 55139 Samsun, Turkey
Abstract:Measurements of microscope stage movements in the x and y directions are of importance for some stereological methods such as the optical disector and optical fractionator. The length of stage movements can be measured with great precision and accuracy using a suitable motorized stage, which is generally a computer-assisted instrument. This type of equipment is generally too expensive for and not readily available in many laboratories. This paper describes a simple method to measure the movements of the microscope stage along the x and y directions, which can be used for purposes such as systematic uniform random sampling. It needs a microscope attachment consisting of two dial indicators; one of them is used to measure the amount of stage movement along the x -axis and the other measures the amount of movement along the y -axis. Movements of the stage on the micrometre-scale can be measured easily using this device.
Keywords:Dial indicator  microscope stage  stereology  systematic uniform random sampling              x and y axes