Sampling wand for an ion trap mass spectrometer |
| |
Authors: | Hou Keyong Xu Wei Xu Jian Cooks R Graham Ouyang Zheng |
| |
Affiliation: | Department of Chemistry, ?Weldon School of Biomedical Engineering, and §Center for Analytical Instrumentation Development, Purdue University , West Lafayette, Indiana 47906, United States. |
| |
Abstract: | A new sampling wand concept for ion trap mass spectrometers equipped with discontinuous atmospheric pressure interfaces (DAPI) has been implemented. The ion trap/DAPI combination facilitates the operation of miniature mass spectrometers equipped with ambient ionization sources. However, in the new implementation, instead of transferring ions pneumatically from a distant source, the mass analyzer and DAPI are separated from the main body of the mass spectrometer and installed at the end of a 1.2 m long wand. During ion introduction, ions are captured in the ion trap while the gas in which they are contained passes through the probe and is pumped away. The larger vacuum volume due to the extended wand improves the mass analysis sensitivity. The wand was tested using a modified hand-held ion trap mass spectrometer without additional power or pumping being required. Improved sensitivity was obtained as demonstrated with nano-electrospray ionization (ESI), atmospheric pressure chemical ionization (APCI), and low temperature plasma (LTP) probe analysis of liquid, gaseous, and solid samples, respectively. |
| |