Real-time digital median frequency estimator for surfacemyoelectric signals
Authors:Pratt  AJ Gander  RE Brandell  BR
Affiliation:Division of Biomedical Engineering, University of Saskatchewan, Saskatoon, Canada.
Abstract:A prototype instrument is described that uses a digital signal processor to estimate the median frequency of the myoelectric signal in real time. A set of tenth-order digital low-pass filters is used to track the median frequency. An estimate of frequency is obtained once every 256 ms with a resolution of 1 Hz.
Keywords: