首页 | 本学科首页   官方微博 | 高级检索  
     

双波长法、三波长法在六价铬盲样考核分析中的应用
引用本文:梁鹏山,俞卫峰,汤晓兰.双波长法、三波长法在六价铬盲样考核分析中的应用[J].化学工程师,2014(3):24-26.
作者姓名:梁鹏山  俞卫峰  汤晓兰
作者单位:杭州市余杭区质量计量监测中心,浙江杭州311100
摘    要:采用两种双波长法和一种三波长法对六价铬盲样进行分析,测定波长均为540nm,两种双波长法的参比波长分别为750和660nm,三波长法的参比波长为460和660nm。与单波长法相比,两种双波长法和三波长法有利于提高六价铬标准样品及盲样测定结果的准确度和精密度。

关 键 词:六价铬  二苯碳酰二肼分光光度法  双波长分光光度法  三波长分光光度法

Determination of chromium (Ⅵ) in blind sample by two dual-wavelength spectrophotometric methods and triplex-wavelength spectrophotometric method
LIANG Peng-shan,YU Wei-feng,TANG Xiao-lan.Determination of chromium (Ⅵ) in blind sample by two dual-wavelength spectrophotometric methods and triplex-wavelength spectrophotometric method[J].Chemical Engineer,2014(3):24-26.
Authors:LIANG Peng-shan  YU Wei-feng  TANG Xiao-lan
Affiliation:(Yuhang District Quality Monitoring Center, Hangzhou 311100, China)
Abstract:Two dual-wavelength spectrophotometric methods and triplex-wavelength speetrophotometnc method using diphenylcarbazide as ehromogenic reagen were established to determinate chromium ( VI ) in blind sample.The measuring wavelength is 540nm, and the reference wavelength is 750nm or 660nm in dual-wavelength spectrophotometric method, while in triplex-wavelength spectrophotometric method reference wavelengths are 460nm and 660nm. Compare with single-wavelength spectrophotometric method, dual-wavelength spectrophoto- metric methods and triplex-wavelength spectrophotometric method had better precision and accuracy.
Keywords:chromium( VI )  diphenylcarbazide spectrophotometric  dual-wavelength spectrophotometry  triplexwavelength spectrophotometry
本文献已被 CNKI 维普 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号