Monitoring of the ballistocardiogram with the static chargesensitive bed
Authors:Jansen  BH Larson  BH Shankar  K
Affiliation:Department of Electrical Engineering, University of Houston, TX 77204.
Abstract:The static charge sensitive bed (SCSB) consists of a sensitive movement detector embedded in a mattress. When a subject rests on the bed, a single electrical signal, containing components reflective of cardiac, respiratory, and body movement related motion, is produced. This paper describes a digital signal processing technique to separate the BCG from the SCSB signal. An evaluation of this algorithm was conducted using recordings from normal volunteers. Comparisons with simultaneously recorded reference signals indicated that the algorithm performed satisfactorily in a laboratory environment.