首页 | 本学科首页   官方微博 | 高级检索  
     


An amperometric glucose biosensor based on a screen-printed electrode and Os-complex mediator for flow injection analysis
Authors:Jian Liu  Shanhui Sun  Changchun Liu  Shoushui Wei
Affiliation:aSchool of Control Science and Engineering, Shandong University, Jinan 250061, China;bJinan Central Hospital Affiliated to Shandong University, Jinan 250013, China
Abstract:An amperometric Os-complex mediated glucose biosensor based on a three electrodes planar configuration was constructed using screen printing technique for flow injection analysis. The electrochemical characteristic of the amperometric biosensor was evaluated by the cyclic voltammetry in the absence and presence of Os-complex mediator. The Os-complex mediated biosensor was found to have a CV curve close to an ideal. The applied potential of the biosensor had been explored and −0.1 V vs. Ag|AgCl reference electrode was selected for avoiding excessive interference from other electroactive species. The biosensor also indicated good stability of more than 90% of its original activity for about 14 months stored at 4 °C. The biosensor on the detection of glucose showed a linear response range of 0.5–30 mM and the correlation coefficient up to 0.99 together with a FIA system. The coefficient of variation was 3.3% for injections with a single biosensor (n = 12). The results show that the biosensor has fast response, good sensitivity and reproducibility for flow injection analysis.
Keywords:Biosensor  Screen-printed  Os-complex mediated  FIA  Amperometric
本文献已被 ScienceDirect 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号