首页 | 本学科首页   官方微博 | 高级检索  
     


Development of a blood cardioplegia delivery system for children
Authors:H Watanabe  H Miyamura  M Sugawara  Y Takahashi  S Tatebe  M Shinonaga  M Takahashi  M Haga  M Hiratsuka  S Eguchi
Affiliation:Second Department of Surgery, Niigata University School of Medicine, Japan.
Abstract:We have developed a blood cardioplegia delivery system for children. Essential points of a delivery system in pediatric cardiac surgery are (1) a small amount of priming volume of a delivery system, and (2) slow, steady infusion of a cardioplegic solution. We changed a heat exchanger to a smaller one for reduction of priming volume, and changed a roller pump tube to a smaller one for slow, steady infusion. Thus, priming volume of a delivery system has reduced from 180 to 100 ml, and we can infuse a cardioplegic solution at a steady rate less than 10 ml/min. Our clinical experience with this system suggests that this blood cardioplegia delivery system is useful for pediatric cardiac surgery.