地质样品中微量元素的高效测试方法
引用本文:王晓媛,殷学博,曾志刚,杨宝菊,陈帅. 地质样品中微量元素的高效测试方法[J]. 质谱学报, 2014, 35(1): 24-31. DOI: 10.7538/zpxb.youxian.2013.34.0006
作者姓名:王晓媛  殷学博  曾志刚  杨宝菊  陈帅
作者单位:1. 中国科学院海洋研究所海洋地质与环境重点实验室,山东 青岛266071;2. 中国海洋大学海洋地球科学学院,山东 青岛266100;3. 中国科学院大学,北京100049
基金项目:国家重点基础研究发展计划(973计划)(2013CB429703);国际海域资源调查与开发“十二五”项目(DY125-12R-02和DY125-11-R-05);山东省自然科学杰出青年基金项目(JQ200913)资助
摘    要:设计并加工了具有双内弧密封结构的防腐高效溶样罐,在消解地质样品时,加入内标103Rh校正样品中元素的浓度,以内标185Re校正分析信号的漂移,使用电感耦合等离子体质谱测定深海沉积物标准物质和国际地质标准物质中的34种微量元素浓度。与以往消解方法相比,该方法密封性好、酸消耗少,只需1.5 mL浓HF、0.5 mL浓HNO3和0.25 mL HClO4就可将40 mg地质样品在短时间内消解完全,消解效率高。前处理过程中加入内标103Rh,样品稀释时无需准确定容,节省操作时间。方法检测下限为0.001~0.621 μg/g,精密度(RSD,n=10)小于5%,准确度小于6.5%(其中大多数元素小于3%),适用于多种地质样品中多元素的分析。

关 键 词:高效消解  内标  微量元素  地质样品  

High Efficiency Determination of Trace Elements in the Geological Samples
WANG Xiao-yuan,YIN Xue-bo,ZENG Zhi-gang,YANG Bao-ju,CHEN Shuai. High Efficiency Determination of Trace Elements in the Geological Samples[J]. Journal of Chinese Mass Spectrometry Society, 2014, 35(1): 24-31. DOI: 10.7538/zpxb.youxian.2013.34.0006
Authors:WANG Xiao-yuan  YIN Xue-bo  ZENG Zhi-gang  YANG Bao-ju  CHEN Shuai
Affiliation:1. Key Laboratory of Marine Geology and Environment, Institute of Oceanology, Chinese Academy of Sciences, Qingdao 266071, China;2. College of Marine Geosciences, Ocean University of China, Qingdao 266100, China;3. University of Chinese Academy of Sciences, Beijing 100049, China
Abstract:A method for determining 34 trace elements in deep-sea marine sediments and international geological samples by ICP-MS was proposed. Samples were digested in anticorrosive efficient digestion vessel, which had a double inner arc seal design. The internal standard 103Rh was used to calibrate the trace element concentrations, and the internal standard 185Re was selected to compensate the drift of analytical signals. Compared with previous digestion methods, this method used less acid in a good seal condition, only 1.5 mL HF, 0.5 mL HNO3 and 0.25 mL HClO4 can digest 40 mg geological samples completely in a short time. Because the internal standard 103Rh was added during the pre-treatment process, so there was no need to dilute the digested sample to an accurate volume, and thus saved a lot of time. This method detection limits are 0.001-0.621 μg/g, the determination precisions (RSD, n=10) are less than 6.5%, and the relative errors between this work and the recommend values are less than 5%, in which most are less than 3%. This method can meet the requirements for analyzing a large number of trace elements in many geological samples.
Keywords:high-efficiency digestion  internal standard  trace element  geological sample  