红外辐射温度计专用数字表的研制及评价
引用本文:盛山菊,卢小丰,唐小洁,宦克为,王雪莲.红外辐射温度计专用数字表的研制及评价[J].计量学报,2019,40(1):20-24.
作者姓名:盛山菊  卢小丰  唐小洁  宦克为  王雪莲
作者单位:长春理工大学,吉林长春,130022;中国计量科学研究院,北京,100029;西安工程大学,陕西西安,710048
基金项目:科技部NQI专项(2017YFF0205902)
摘    要:日本CHINO公司生产的IR-RST系列标准辐射温度计由于没有设计温度显示装置,无法实时显示温度,针对此类型温度计设计了一款专用的电压测量和温度显示,同时可提供24V直流电压的电测显示仪表。利用高等级的电压源和数字电压表,修正了本专用数字表的误差系数,并考核了稳定性,得到其最大相对误差不超过0.03%。与一款名义波长为0.65 μm的IR-RST辐射温度计配合使用,通过TG HT-9500型高温炉对其组合进行内插分度,实验结果证实分度后校准准确度优于0.03%,1 300 ℃时系统测量结果的扩展不确定度为0.62 ℃。

关 键 词:计量学  红外辐射温度计  电测显示仪表  校准  不确定度
收稿时间:2018-08-30

Development and Evaluation of Digital Meter for an Infrared Radiation Thermometer
SHENG Shan-ju,LU Xiao-feng,TANG Xiao-jie,HUAN Ke-wei,WANG Xue-lian.Development and Evaluation of Digital Meter for an Infrared Radiation Thermometer[J].Acta Metrologica Sinica,2019,40(1):20-24.
Authors:SHENG Shan-ju  LU Xiao-feng  TANG Xiao-jie  HUAN Ke-wei  WANG Xue-lian
Affiliation:1. Changchun University of Science & Technology, Changchun, Jilin 130022, China
2. National Institute of Metrology, Beijing 100029, China
3. Xi‘an Polytechnic University, Xi'an, Shaanxi 710048, China
Abstract:The IR-RST standard radiation thermometers manufactured by CHINO Company, Limited, Japan, which can not display the temperature in real-time because the display unit hadn’t been designed. Therefore, an electrical measurement display device was designed which can measure the voltage, display the temperature and supply 24V DC voltage for this kind of infrared radiation thermometer. Calibrated a against high-level voltage power source and a voltmeter, the correction coefficients of the display instrument were got, the stabilitywas evaluated as well, the maximum relative error of voltage measurement can be achieved with in 0.03%. Assembled with a 0.65μm IR-RST radiation thermometer, it was calibrated against with a TG HT-9500 type high-temperature furnace as the source, the accuracy was better than 0.03% confirmed by the experimental results, and the extended uncertainty of the calibration system was 0.62℃ at 1 300℃.
Keywords:metrology  infrared radiation thermometer  digital display instrument  calibration  uncertainty  