Effect of measurement duration on accuracy of pulse-counting
Abstract:In this study, the relation between the measurement duration and accuracy of pulse-counting was quantitatively examined with special reference to low-frequency fluctuations in heart rate variability. The interbeat intervals of 70 healthy male subjects were measured in standing, sitting and supine positions. Pulse rates for various durations were calculated by objective-scoring simulation based on the heartbeat recordings of the subjects. The duration of pulse-counting continuously varied from 6 to 60 s in the simulation. Simulated pulse rates were compared with the rate calculated from the 60 s that includes the given duration, and the absolute difference between the two rates was defined as the error. Average errors of pulse-counting for 15 s were 1.89, 1.89 and 1.80 bpm for standing, sitting and supine positions, respectively. No difference in error was observed between standing and sitting positions; however, smaller errors were observed in the supine position.
Keywords:pulse  count  accuracy  error  heart rate variability