首页 | 本学科首页   官方微博 | 高级检索  
     


FEM analysis of predicting electrode-myocardium contact from RF cardiac catheter ablation system impedance
Authors:Cao Hong  Speidel Michael A  Tsai Jang-Zern  Van Lysel Michael S  Vorperian Vicken R  Webster John G
Affiliation:Department of Electrical and Computer Engineering, University of Wisconsin, Madison 53706, USA.
Abstract:We used the finite-element method (FEM) to model and analyze the resistance between the catheter tip electrode and the dispersive electrode during radio-frequency cardiac catheter ablation for the prediction of myocardium-electrode contact. We included deformation of the myocardial surface to achieve accurate modeling. For perpendicular catheter contact, we measured the side view of myocardial deformation using X-ray projection imaging. We averaged the deformation contour from nine samples, and then incorporated the contour information into our FEM model. We measured the resistivity of the bovine myocardium using the four-electrode method, and then calculated the resistance change as the catheter penetrated into the myocardium. The FEM result of resistance versus catheter penetration depth matches well with our experimental data.
Keywords:
本文献已被 PubMed 等数据库收录!