首页 | 本学科首页   官方微博 | 高级检索  
     


Performance of the Coumatrak system in a large anticoagulation clinic
Authors:PR Foulis  PM Wallach  HM Adelman  BH Sanford  J McCain  D Reed  CM Schlede  CU Kokseng  CD Taylor
Affiliation:University of South Florida Health Sciences Center, Tampa.
Abstract:There is a need to monitor anticoagulation accurately, inexpensively, and rapidly. The accuracy and precision of a simple fingerstick method was studied in a large outpatient anticoagulation clinic using the Coumatrak method. The Coumatrak apparatus has been studied in the home setting, and three recent reports suggest that it is practical, accurate, and possibly superior to the standard method. These results differ from recently published studies. This technique was found to be less than acceptable in precision and accuracy. This method requires further study before it can be recommended for wide-spread use in making decisions for patient care.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏