Mechanical failure of the intramedullary hip screw in a subtrochanteric femoral fracture |
| |
Authors: | J Rantanen H Aro |
| |
Affiliation: | Department of Surgery, University of Turku, Finland. |
| |
Abstract: | The Gamma nail, an implant specifically designed for intertrochanteric and subtrochanteric femoral fractures, has been criticized for its high risk of secondary shaft fractures. A modified design, the intramedullary hip screw has recently been introduced to correct this complication. We present a case of mechanical failure of this new implant that occurred in a pathological subtrochanteric fracture. The centering sleeve of the implant became loose and migrated while the head screw penetrated the acetabulum. Refixation was successful. |
| |