首页 | 本学科首页   官方微博 | 高级检索  
     


Mechanical failure of the intramedullary hip screw in a subtrochanteric femoral fracture
Authors:J Rantanen  H Aro
Affiliation:Department of Surgery, University of Turku, Finland.
Abstract:The Gamma nail, an implant specifically designed for intertrochanteric and subtrochanteric femoral fractures, has been criticized for its high risk of secondary shaft fractures. A modified design, the intramedullary hip screw has recently been introduced to correct this complication. We present a case of mechanical failure of this new implant that occurred in a pathological subtrochanteric fracture. The centering sleeve of the implant became loose and migrated while the head screw penetrated the acetabulum. Refixation was successful.
Keywords:
本文献已被 PubMed 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏