首页 | 本学科首页   官方微博 | 高级检索  
     


(HP)Ge Measurement of spectra for diagnostic X-ray beams
Authors:Nogueira M S  Mota H C  Campos L L
Affiliation:Centro de Desenvolvimento da Tecnologia Nuclear, R. Mário Werneck, S/N, 30123-970, Belo Horizonte, MG, Brazil. mnogue@cdtn.br
Abstract:The X-ray spectra applied in diagnostic radiology to the energy range between 50 and 125 kVp, were experimentally determined using a high-purity germanium detector, (HP)Ge, coupled to a multichannel analyser. A simple stripping procedure was implemented, according to that described by Seelentag and Panzer, to correct the pulse height distribution and then to determine the photon spectra. The measurements were performed for those conditions where the X-ray beam is transmitted through Pb filters 0.2-0.7 mm thick. It was necessary to estimate the effective dose in the same way that it is done for radiology diagnosis room shielding projects. The spectra were also determined to the X-ray qualities recommended by the IEC for primary diagnostic X-ray and for X-ray beams attenuated by aluminium filters 2.5-42.5 mm thick. The spectra obtained were compared with data derived from the literature presenting good agreement.
Keywords:
本文献已被 PubMed Oxford 等数据库收录!
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司  京ICP备09084417号